Chapter |
Connection Specifications
Confidential, limited rights
PN AA83865 | Revision A | 25-Oct-2019
Instruction Notice, ARL EQUINOX 5000-5500
Page
29
5.2 Cooling Fluid
The cooling circuit of the ARL EQUINOX 5000/5500 has to be connected to an external
cooling device. The cooling specifications are given in the table below.
Parameter
Min
Max
Critical values for
generator safety stop
Flow rate
3.5 L/min
6 L/min
< 3 L/min
Pressure
3 bar
8 bar
Temperature*
18 °C
25 °C
< 10 °C
> 35 °C
Conductivity
10 μS/cm
250 μS/cm
Harness (German degrees)*
1 D
4 D
Acidity (pH)*
7.5
9.5
Input/Output
Quick fittings for polyurethanes
hoses. Øinner 7 mm - Øouter 10 mm
* Taken from X-Ray tube manufacturer specification.
The cooling fluid is typically composed by < 10% of ethylene glycol in deionized water.