Thermo Scientific AquaSensors
™
AnalogPlus
™
Contacting Conductivity Sensor User Guide
9
3.2.
Mounting Hardware
The Thermo Scientific AquaSensors 1.0 inch Contacting Conductivity sensor can be
installed in standard plastic and metal 1-inch full bore tees, union mount tees, immersion
mounts and sanitary mounts.
K 1.0
Save the protective cap and use it
whenever the sensor is removed
from service.
DO NOT USE PIPE SEALANT.
Pipe sealants may not provide
adequate sealing or may react with
different plastic materials. Use thread
tape.
If the sensor does not calibrate, refer
to Section 4 to review maintenance,
refurbishment and troubleshooting
options.
DO NOT OVER TIGHTEN!
Maximum torque: 65 lbs
⋅
inch (7 N
⋅
m)
The sensor should be positioned with
the electrodes into the flow. This will
insure that entrained air does not
form a pocket around the sensor
head.
K 0.1
Figure 3.2:
1 Inch Sensor Body