Detailed Pipe Tee Installation Instructions
1.
Wiring:
a.
Route the cable either through conduit or on a cable tray.
b.
Insert the sensor cable into the Model C99 controller through a watertight cord grip.
Conduit holes are provided for the cord grip on the bottom of the C99 controller.
c.
Connect the sensor wires into the C99 sensor terminal block as shown in the C99
manual.
2.
Calibrate:
a.
Power the C99 controller.
b.
Remove the protective cap from the sensor head. Connect and calibrate the
Conductivity sensor as explained in the C99 manual. Subsequent calibrations
should be scheduled based on process demands.
c.
If measurement response time is slow or if the sensor will not calibrate, refer to
Section 4 for maintenance, refurbishment and troubleshooting suggestions.
3.
Mount:
a.
Apply Teflon
®
tape to the sensor body threads.
b.
Insert the sensor into the tee and slowly turn clockwise until secure. Tighten the
sensor with a wrench until snug to prevent leaking. Be sure the pipe remains full
when the sensors are installed.