- 7 -
2.3.3
Connecting pH/ORP Electrode
1. If the pH/ORP electrode has a BNC connector, remove the BNC
connector from the cable.
NOTE:
Oakton Instruments offers an optional ‘BNC to Spade Lug
adapter’ (Order code: 05994-90) that can be used with pH/ORP
electrode without removing the BNC connector.
2. Strip the insulation of the cable so that the bare wires are exposed
enough for connection as shown in Figure 3.
Figure 3 : pH/ORP Electrode Cable
NOTE:
Make sure to strip inner black layer (screen) to expose the
clear sheath.
3. Connect pH sensing wire to
Pin 1
of J204 connector
4. Connect pH reference wire to
Pin 2
of J204 connector
Insulation
pH Referenc
Clear sheath
pH Sensing W
Screen
e Wire
ire