2.3 Connecting Peripherals
2.3.1
Connection Terminals
Remove Back Cover
:
•
Remove the screws from the four corners at the back of the pH
Transmitter.
•
Remove the back cover. The connectors are exposed on the back
PCBA as shown in the Figure 1 below.
Connectors
:
•
J202 – 9V DC power
•
J204 - pH electrode & Temperature probe connections (wiring has to
be done in the detachable connector:
1. Pin 1
: pH Sensor
2. Pin 2
: pH Reference
3. Pin 3
: No Connection
4. Pin 4
: No Connection
5. Temp I/P
: Pt 100 Compensate
6. Temp Sens
: Pt 100 Sense
7. Temp Gnd
: Pt 100 GND
8. PMP
: Potential Matching Pin
Figure 1: Outer Side of Back PCBA
J204
J202
Pin 1
Pin 2
Pin 3
Pin 4
Temp I/P
Temp Sens
Temp Gnd
PMP