3.3.1 Mounting the Diverter Valve Assembly
Mounting the diverter valve assembly on the fraction collector arm is the first major task.
The diverter valve assembly attaches to the fraction collector arm using two knurled
finger-tight screws.
1.
Place the diverter valve assembly on the fraction collector arm and mount it with two
knurled finger-tight screws. No tools are required to fasten the screws. Fig. 3 shows
the assembly attached.
Fig. 3: Diverter valve assembly
2.
Attach the mounting block to the AFC-3000 rear panel with the finger-tight mounting
block screws provided (
→
Fig. 4: Attaching the mounting block
3.
Connect the diverter valve cable to the auxiliary I/O connection on the AFC-3000 rear
panel (
→
Fig. 2, page 18). Tighten the knurled nut finger-tight.
4.
Screw the PEEK drop former finger-tight into the diverter valve.
Knurled screws
Содержание AFC-3000
Страница 2: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Operating Instructions ...
Страница 4: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page II Operating Instructions ...
Страница 18: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 12 Operating Instructions ...
Страница 28: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 22 Operating Instructions ...
Страница 48: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 42 Operating Instructions ...
Страница 76: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 70 Operating Instructions ...
Страница 78: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 72 Operating Instructions ...
Страница 82: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 76 Operating Instructions ...
Страница 86: ...UltiMate 3000 Series AFC 3000 Automated Fraction Collector Page 80 Operating Instructions Sample run ends here ...