4
Instrument Driver Configuration
Foundation Instrument Configuration
Thermo Scientific
Accela 600 and 1250 Pump Hardware Manual
73
6. In the Configured Devices area, double-click the
Accela 600
or the
Accela 1250
icon.
The Accela
pump type
dialog box appears (see
). The Serial Number list contains
the serial number of the Accela pump that is powered on and connected to the data
system computer.
7. Ensure that the Serial Number list displays the pump’s serial number and that the Status
readback displays Connected.
The serial number is listed on the pump’s back panel (see
).
Figure 63.
Accela
pump type
dialog box
8. If your Accela pump has a leak sensor, select the
Leak Sensor
Installed
check box.
The leak sensor is an option for the Accela 600 Pump and the Accela 1250 Pump. The
leak sensor has an optical sensor that turns red when covered by liquid.
For instructions about installing the leak sensor, see
. For information about setting up leak sensor actions, such as stopping a run, see
“Specifying the Leak Sensor Settings in the Xcalibur Data System”
9. Click
OK
to accept the configuration settings and close the Accela
pump type
dialog box.
10. Click
Done
to close the Instrument Configuration application.
If you have a dual-pump system, continue to the next procedure.
Note
The firmware version compatible with the current version of LC Devices is
subject to change.