Hardware Upgrades
Leak Sensor Installation
Thermo Scientific
Accela 600 and 1250 Pump Hardware Manual
53
Leak Sensor Installation
Thermo Fisher Scientific recommends installing the optional leak sensor to stop the solvent
flow from the pump when a leak occurs.
The Leak Sensor Kit (P/N 00960-01-00024) contains these items:
• Leak sensor with attached cable
• Mounting bracket
• Locating fixture
• Two-component adhesive
To install the optional leak sensor
1. Turn off the power to the pump and unplug the pump from line power.
2. Glue the mounting bracket for the leak sensor to the pump’s drip tray as follows:
a. Place the locating fixture against the front panel of the pump as shown in
Figure 41.
Appropriate position for the locating fixture
b. Mix the two-component adhesive on a disposable surface, such as a sheet of weighing
paper.
c.
Add a thin layer of adhesive to the back side of the mounting bracket.
Locating fixture