3
Fluoride Ion Selective Electrode User Guide
Required Equipment
1. Thermo
Scientifi c Orion ISE meter, such as the Orion 4-Star
pH/ISE meter, Orion 5-Star pH/ISE/DO/conductivity meter
or Orion DUAL STAR meter.
Fluoride electrodes can be used on any ISE meter with
a BNC connection. The electrodes can also be used on
meters with a variety of inputs when an adapter cable is
used. Visit www.thermoscientifi c.com/water for details.
2. Thermo
Scientifi c Orion fl uoride ion selective electrode.
The 9409BN and 9409SC fl uoride half-cell electrodes
require a separate reference electrode,
Cat. No. 900100.
3. Magnetic stirrer or Orion stirrer probe, Cat. No. 096019.
The Orion stirrer probe can be used with 3-Star, 4-Star, 5-
Star benchtop meters, and Dual Star meter.
4. Volumetric
fl asks, graduated cylinders and beakers.
Plastic labware is highly recommended for fl uoride analysis.
5. Distilled or deionized water.
6. Fluoride electrode fi lling solution.
Use Optimum Results A fi lling solution, Cat. No. 900061,
for the 9609BNWP fl uoride electrodes.
Use single junction reference electrode fi lling solution,
Cat. No. 900001, for the reference electrode used with the
9409BN and 9409SC fl uoride electrodes.
7.
Fluoride calibration standards.
Cat. No.
Description
940906
0.1 M NaF, 475 mL bottle
940907
100 ppm F¯, 475 mL bottle
040908
10 ppm F¯ with TISAB II, 475 mL bottle
040907
2 ppm F¯ with TISAB II, 475 mL bottle
040906
1 ppm F¯ with TISAB II, 475 mL bottle