9
Fluoride Ion Selective Electrode User Guide
Electrode Storage
9409BN and 9409SC Fluoride Half-Cell Electrode
The fl uoride half-cell electrodes should be rinsed thoroughly
with distilled water and stored dry in the air at all times. When
storing the electrode for long periods of time, cover the sensing
element with the protective shipping cap.
900100 Single Junction Reference Electrode
The single junction reference electrode may be stored in the
single junction electrode fi lling solution, Cat. No. 900001,
between sample measurements and up to one week. A 0.01 M
or 100 ppm fl uoride standard is also an acceptable storage
solution. The fi lling solution inside the electrode should not be
allowed to evaporate, as crystallization will result.
For storage longer than one week, drain the reference electrode,
fl ush the inside with distilled water and store the electrode dry.
9609BNWP Fluoride Combination Electrode
For storage between measurements and up to one week, store
the electrode in electrode fi ll solution, Cat. No. 900061, with
fl uoride added. The fl uoride concentration of the storage solution
should be close to least concentrated fl uoride calibration
standard. Do not add TISAB to the storage solution. The fi lling
solution should not be allowed to evaporate, as crystallization
will result. If fi ll solution is unavailable, store in 2 M potassium
nitrate with fl uoride added.
For storage longer than one week, drain the electrode. Flush
the chamber with distilled water. Disassemble the electrode as
described in the electrode maintenance section. Rinse the inner
sleeve with distilled water and blot dry with a Kimwipe. Do not
touch the reference element other than to blot dry. Let air dry
and then reassemble for storage. Use the protective shipping
cap to cover the sensing element.