![Thermo Scientific 8120 Скачать руководство пользователя страница 28](http://html.mh-extra.com/html/thermo-scientific/8120/8120_operating-and-maintenance-manual_1104928028.webp)
Section 6
Service and Maintenance
This section contains the information regarding the liquid cylinder care
and maintenance. It includes the particular maintenance procedures for
changes to the operating pressure, service pressure and liquid service
changes. When performing a procedure that is described in this section,
refer to the Operation section for a components item number and
location.
Before implementing any procedure described in this section, it is
recommended that the Safety section and Product Safety Bulletins be read
and fully understood.
Always keep cylinders clean and free from grease and oil.
When repairing containers, use only parts which are considered compatible
with liquid to be used and which have been properly cleaned for this
specific liquid service.
Do not use regulators, fittings, or hoses which were previous used in a
compressed air service. Use only compatible sealants or Teflon tape on the
threaded fittings. All new joints should be leak tested with an specifically
compatible leak test solution.
Caution
Before conducting maintenance or replacing parts on a cylinder,
release container pressure in a safe manner. Replacement of certain cylinder
parts may require that the container contents be completely emptied.
The Cryo-Cyl Series liquid cylinders are designed to hold any of the gas
products specified. They can easily be modified to work as well with
nitrogen as oxygen. The fittings and decals need to be changed and the
inner vessel needs to be purged.
If a cylinder is changed from inert (argon or nitrogen) to CO
2
service, the
relief valve must be changed to a CO
2
relief valve.
Caution
Once a cylinder is used in CO
2
service, it can not be used for
other gas products, especially oxygen or nitrous oxide.
Caution
Whenever converting a Nitrogen or Argon cylinder to Oxygen
use, inspect the cylinder to assure cleanliness.
180 liter LN2 Supply Tank 6-1
Thermo Scientific
Safety
Cleaning
Changing Service