Thermo Scientific Barnstead LabTower AFT
Pure water system
12
Electrical connections / external switched mode power supply AFT 20
Input voltage
AC 100 – 240 V, 50 – 60 Hz, 5 – 3.8 A
Output voltage
DC 24 V, 3.8 A
System connection
DC 24 V, 80 W
Serial interface
RS 232
Protection Class
Class
II
(external SMPS certified as Class
I
)
Electrical connections / external switched mode power supply AFT 40
Input voltage
AC 100 – 250 V, 50 – 60 Hz, 4 – 2.5 A
Output voltage
DC 48 V, 2.5 A
System connection
DC 48 V, 120 W
Serial interface
RS 232
Protection Class
Class
II
(external SMPS certified as Class
I
)
Airborne sound emission
Sound-pressure level
49 db(A)
Ambient conditions (DIN EN 61010-1 (VDE 0411-1):2011-02)
Usage
Indoor rooms
Height
Up to 2000 m
Temperature range
From 5° C to 40° C
Relative humidity
Maximum relative humidity 80 % at temperatures of
up to 31° C, linearly decreasing to 50 % relative hu-
midity at 40° C
Line-voltage variation
Not more than
±
10 % of the line voltage
Transient overvoltages
As usually occur in the supply network (overvoltage
category II acc. To IEC 60364-4-443).
Note: The rated level of transient overvoltage is the
withstand impulse voltage acc. To overvoltage cate-
gory II of IEC 60364-4-443
Ventilation requirements
There are no special requirements with regard to ven-
tilation.
Degree of pollution
2
Materials of parts that contact water
Pump head
Nylon with glass fibre
Filter cartridge
PP
Rinsing solenoid valve
PA
Conductivity measuring cell
POM, stainless steel
Distribution block
POM
Connectors
POM
Hoses
PE
Gaskets
EPDM