C-Link Protocol Commands
Measurement Configuration
Thermo Fisher Scientific
Model 49
i
Instruction Manual
B-29
Measurement
Configuration
range
high range
low range
These commands report the O
3
range in single range mode, or the high and
low ranges in dual or auto range mode, according to
. If the mode
is incorrect, the instrument responds with “can’t, wrong settings”. The
example that follows reports that the O
3
range is 50 ppb.
Send:
range o3
Receive:
range o3 0: 5000E-2 ppb
set range
selection
set high range
selection
set low range
selection
These commands select the O
3
full-scale range, according to
example that follows sets the O
3
full-scale to 2,000 ppb (2 ppm).
Send:
set range o3 5
Receive:
set range o3 5 ok
Table B-6.
Range Settings
Code
ppm
mg/m
3
0
0.05
0.1
1
0.1
0.2
2
0.2
0.4
3
0.5
1
4
1
2
5
2
4
6
5
10
7
10
20
8
20
40
9
50
100
10
100
200
11
200
400
12
C1
C1
13
C2
C2
14
C3
C3
Содержание 49i
Страница 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Страница 4: ......
Страница 6: ......
Страница 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Страница 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 162: ...Preventive Maintenance Ozone Scrubber Test 5 12 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...
Страница 244: ...Optional Equipment Mounting Options 9 6 Model 49i Instruction Manual Thermo Fisher Scientific Figure 9 2 Bench Mounting ...