Maintenance
When the Finnpipette Novus is not in use, make sure it is stored in an upright position. We
recommend a Finnpipette stand for this purpose.
The part # refer to exploded views beginning at page 93.
Daily checking
The pipette should be checked at the beginning of each day for dust and dirt on the outside
surfaces of the pipette. Particular attention should be paid to the tip cone. No other solvents
except 70 % ethanol should be used to clean the pipette.
Short-term service
If the pipette is used daily it should be checked and lubricated every three months. The servicing
procedure starts with choosing the Service-mode from the menu (Menu -> Options -> Service).
Pipettes up to 1000 µl
1. Remove the lower tip ejector sleeve 13 by pulling it out (snap fitting).
2. Push down the ejector button and take the upper ejector top 11 between fingers.
3. Release the ejector button to up position, but hold the tip ejector top 11
in down position with the fingers.
4. Remove the tip cone by pulling it out (snap fitting).
5. Press and hold the trigger down to drive out the piston.
6. Insert the special piston removal tool and pull out the piston 14. The number
2 end is for 300µl and 1000µl pistons, the number 1 end for all the smaller
volumes.
See picture.
Release the trigger.
7. Check the tip cone for foreign particles.
8. Grease the cleaned parts with the lubricant that comes with the pipette.
9. Reassemble the pipette components.
10. Insert carefully the piston to the tip cone.
11. Push the tip cone back to the handle while holding
the ejector top 11 down, until you hear a “click”.
12. Press trigger to reconnect the piston to drive mechanism.
13. Press READY to return from service mode.
5ml and 10ml pipettes
1. Pull down the lower tip ejector sleeve.
2. Note that the tip ejector button is in fully up position.
3. Remove the tip cone by pulling it out (snap fitting).
4. Remove the ejector sleeve by pulling it to two parts
(snap fitting).
5. Push the snap joints to release and remove the cylinder 14.
6. Push forward and clean the piston with a dry napless cloth.
7. Check the tip cone for foreign particles.
8. Grease the cleaned parts with the lubricant that comes
with the pipette.
9. Reassemble the pipette components in reverse order.
10. Push the tip cone back to the handle while holding the ejector top 11 down, until you hear
a “click”
11. Press trigger to reconnect the piston to drive mechanism.
12. Press READY to return from service mode.
Long-term service
If the pipette is used daily it should be serviced every six months. The servicing procedure starts
with the disassembly of the pipette.
Pipettes up to 1000 µl
1. Remove the lower tip ejector sleeve 13 by pulling it out (snap fitting).
2. Push down the ejector button and take the upper ejector top 11 between fingers.
3. Release the ejector button to up position, but hold the tip ejector top 11 in down position
with the fingers.
4. Remove the tip cone by pulling it out (snap fitting).
5. Press and hold the trigger down to drive out the piston.
English
1
2
1
2