Operation
Settings
Thermo Scientific
43iQ Instruction Manual
3-89
The Alarm Setpoints screen allows the user to view and set all settable
alarm minimum and maximum values.
Note
User cannot set alarm limits outside of the acceptable range. The
minimum and maximum alarm limit can also be set by pressing on the
corresponding buttons located in the Settings>Health Check>Status and
Alarms screen. See “
▲
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Settings>Instrument Settings>Alarm Setpoints
The Alarm Setpoints screen contains the following information:
●
Across:
●
Variable:
Lists the items that have settable alarm limits.
●
Value:
Displays the current value for each item.
●
Low Alarm:
User sets low alarm for item.
●
High Alarm:
User sets high alarm for item.
●
Units:
Units for each item (not settable).
●
Down:
●
SO2 Concentration:
SO
2
concentration alarm.
●
Bkg Check Offset:
User can set the maximum allowable background
reading offset for calibration and calibration checks. This is set with
the high alarm only.
Alarm Setpoints
Содержание 43iQ
Страница 1: ...43iQ Instruction Manual Sulfur Dioxide Analyzer 117568 00 1Sep2021 ...
Страница 2: ......
Страница 8: ......
Страница 18: ......
Страница 34: ...Operation Instrument Display 3 8 43iQ Instruction Manual Thermo Scientific ...
Страница 162: ......
Страница 166: ...Maintenance Pump Rebuilding 5 4 43iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 170: ......
Страница 180: ......
Страница 188: ...Servicing Fan Replacement 7 8 43iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 228: ......
Страница 238: ......
Страница 262: ......
Страница 270: ......
Страница 282: ......
Страница 283: ......