Installation
Setup Procedure
Thermo Fisher Scientific
Model 146
i
Instruction Manual
2-5
Figure 2-4.
Replacing the Glass Chamber Assembly
h. Tighten knurled screw with finger. Do not use tools to tighten.
i.
Replace oven cover, being careful to place tubing and wire in slot of
cover.
4. Connect the
EXHAUST bulkhead to a suitable vent. The exhaust line
should be 1/4-inch OD with a minimum ID of 1/8-inch. The length of
the exhaust line should be less than 10 feet. Verify that there is no
restriction in this line.
5. Connect a suitable recording device to the rear panel connector. For
detailed information about connecting to the instrument, refer to:
“External Device Connection Components”
“Terminal Block and Cable Kits”
“Instrument Controls Menu” in the Operation chapter
For detailed information about troubleshooting a connection, refer to
.
Permeation
Oven Top
See Expanded View
Expanded View
Chamber Top
Oven Top
Glass
Chamber
Assembly
Permeation
Oven Bottom
See Procedure
Содержание 146i
Страница 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Страница 4: ......
Страница 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 148: ......
Страница 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Страница 272: ...9 14 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 4 Rack Mount Option Assembly ...
Страница 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Страница 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Страница 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...
Страница 276: ...Optional Equipment Mounting Options 9 18 Model 146i Instruction Manual Thermo Fisher Scientific ...