Chapter 2 Installation
2-4
SCRUBBER INSTALLATION
The 1160 are shipped with the adsorbent scrubber systems empty. Before turning the
instrument on, make sure that the canisters contain fresh packing. Please refer to Figures
2-3, 2-4 and 4-1 while reviewing the following procedure.
1.
Loosen the hold down screw, swing the handle aside and remove the top plate.
2.
Insure that both felt-pad filters are flat against the bottom of the plastic
canister.
3.
Fill both the inner and outer cylinders of the canister to approximately ¼” of
the top with the proper packing. Do not over-fill or the cover may not seat
properly.
4.
Replace the top plate, insuring that the O-ring is properly seated. Note that
there are two O-rings in the base plate and one in the scrubber cap, as shown
in Figure 2-3. The O-rings can be identified by comparison with the full size,
or 1:1, drawing included in Figure 2-4.
5.
Re-seat the canisters, being sure that the O-rings are properly seated, and then
swing the handle back up into position.
6.
Tighten the hold down screw “finger-tight.” Then, use a 3/8” open-end
wrench to turn the screw approximately
¾
turn more. Do not over-tighten.
Содержание 1160
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 22: ...Chapter 2 Installation 2 6 Figure 2 4 Scrubber Assembly Component P N 104558 00 P N 104558 00 P N 104557 00 ...
Страница 30: ...Chapter 4 Maintenance 4 2 Figure 4 1 Scrubber Parts List ...
Страница 34: ......
Страница 46: ...C 2 ...
Страница 47: ...C 3 ...
Страница 48: ...C 4 ...
Страница 49: ...C 5 ...
Страница 50: ...C 6 ...
Страница 51: ...C 7 ...
Страница 52: ...C 8 ...