Thermo Scientific 02LREETSA Скачать руководство пользователя страница 7

 

Intended Use

 Statement

 

 

The  Refrigerators/Freezers described  in  this  manual  are  for 

professional use  only.  These  products  are  intended  for  use  in 

research for the storage of samples or inventory in the following 

temperature ranges: 

 

 

 

 

Warning

 

Only  Explosion  Proof  Units  or 

Flammable Material  Storage  Units 

are  to  be  used  for the  storage of 

flammable inventory/samples. 

General Purpose Freezers: 

 -12°C to -24°C 

 
General Purpose Refrigerators:      1°C to 8°C 
 
Explosion proof & FMS Freezers:   -12°C to -24°C 
  

These  are  not  considered  medical  devices  and  have  therefore 

not  been registered with a medical device regulatory body (e.g. 

FDA):  that  is,  it  has  not  been  evaluated  for  the  storage  of 

samples for diagnostic use or for samples to be re-introduced to 

the body. 

 

NOTE: Do not store corrosive materials in these units. Any damage 

which occurs due to storage of corrosives will not be covered under 

warranty claims. 

 

Corrosive Materials Requiring Refrigerated or Frozen Storage 

 Only use models rated as Corrosion Resistant 

 Corrosion Resistant does not mean Corrosion Proof – Care in 

storage is still required 

 Store only corrosive reagents/samples which truly need 

reduced temperature storage. 

 Flammable corrosive materials require Flammable Material Storage 

or Explosion Proof models labeled for the storage of corrosives 

 Containers must be wiped clean of moisture and chemical 

residue before being introduced into the unit and upon return 

from use. 

 Containers must be sealed with either vinyl tape or Parafilm ® 

 Reagents/samples which release HX (X= F, Cl, Br, I) on contact 

with moisture (e.g. Acyl halides, Organosilyl halides etc.) are 

particularly damaging to metals. 

 Volatile amines will react with HX depositing salts which will 

lead to corrosion of metal surfaces. 

 Volatile organic acids can exacerbate metal surfaces 

already compromised, be sure these are securely sealed. 

 Refrigerated compartments are cool or cold areas but they are not 

to dry seal your reagents and samples. 

 Bleach solutions release chlorine gas which can react with 

other volatiles in the cooling chamber or directly attacks metal 

surfaces. 

 Periodically clean the interiors, clean up spills or leaking containers 

 Failure to take precautionary actions may lead to damage not 

covered by warranty claims 
Parafilm ® is a registered trademark of Bemis Company, Oshkosh, 

WI, 

 

 

 

Содержание 02LREETSA

Страница 1: ...Installation and Operation Manual Thermo Scientific General Purpose Flammable Storage and Explosion Proof Freezers...

Страница 2: ...2016 Thermo Fisher Scientific All rights reserved...

Страница 3: ...General Purpose Flammable Material Storage and Explosion Proof Freezers Explosion Proof 05EFEETSA Flammable Material Storage 05FFEETSA General Purpose 05LFEETSA 02LREETSA...

Страница 4: ...d Installation 13 Shipping Carton 13 Unpacking 13 Location 13 Clearance 13 Electrical 14 Be Advised 14 How to Seal Junction Box Conduit with Fiber and Sealing Compound to Help Protect Against Explosio...

Страница 5: ...s I II or III locations as defined by the US National Electrical Code unless otherwise noted Alert Signals ON OFF Safety Alert Important operating instructions To reduce the risk of injury or poor per...

Страница 6: ...refrigerator or freezer Take off doors Leave the shelves in the place so that children may not easily climb inside If the equipment is used in a manner not specified by the manufacturer the protection...

Страница 7: ...in storage is still required Store only corrosive reagents samples which truly need reduced temperature storage Flammable corrosive materials require Flammable Material Storage or Explosion Proof mod...

Страница 8: ...d determined to remain below the flash point of materials found in Class I Groups C and D All models have heavy gauge rigid steel construction with a durable enamel finish Interiors have epoxy enamel...

Страница 9: ...elves Model Freezer Chamber Dimensions H x W x D inches cm Total Vol Cu Ft Exterior Dimensions H x W x D inches cm 05EFEETSA 29 5 x 18 63 x 19 74 9 x 47 32 x 48 2 cm 5 0 33 13 x 21 38 x 30 25 84 15 x...

Страница 10: ...hazardous materials inside the unit These units are ideal for storing cyclopropane ethyl ether ethylene acetone alcohol benzene butane gasoline hexane lacquer solvent vapors naphtha natural gas or pr...

Страница 11: ...oor shelves Model Freezer Chamber Dimensions H x W x D inches cm Total Vol Cu Ft Exterior Dimensions H x W x D inches cm 05FFEETSA 29 5 x 18 63 x 19 74 9 x 47 32 x 48 2 cm 5 0 33 13 x 21 38 x 29 5 84...

Страница 12: ...ical biological materials may require alarm monitoring back up and temperature performance features The need to add these features should be determined by your industrial hygienist or safety officer T...

Страница 13: ...or shelves Model Freezer Chamber Dimensions H x W x D inches cm Total Vol Cu Ft Exterior Dimensions H x W x D inches cm 05LFEETSA 29 5 x 18 63 x 19 74 9 x 47 32 x 48 2 cm 5 0 33 13 x 21 38 x 25 75 84...

Страница 14: ...ator Chamber Dimensions H x W x D inches cm Total Vol Cu Ft Exterior Dimensions H x W x D inches cm 02LREETSA 15 5 x 16 x 14 39 37 x 40 64 x 35 56 cm 1 8 19 0 x 18 63 x 17 25 48 26 x 47 32 x 43 81 cm...

Страница 15: ...unit has been received Do not discard packing materials until all is accounted for The following items are included in the shipment Refrigerator Freezer Operation Manual Inspection Tag Location Place...

Страница 16: ...ted to its own individual outlet Be Advised WARNING UNLESS UNIT IS SPECIFICALLY DESIGNED FOR COMBUSTIBLE OR FLAMMABLE ATMOSPHERES DO NOT USE IN THE PRESENCE OF FLAMMABLE OR COMBUSTIBLE MATERIALS OR EX...

Страница 17: ...vent the very real threat of gas entering the junction box and a resulting serious explosion After the unit wires have been pulled through the vertical conduit the following procedure is required Turn...

Страница 18: ...terial COMPLETELY SURROUNDS THE WIRING from the top to the bottom completely blocking this end of the vertical conduit Insert fiber rope material down into horizontal conduit opening Pressing down fir...

Страница 19: ...l not to trap air bubbles Immediately wipe off any spilled sealing compound Screw conduit domed cover back onto conduit opening NOTE INITIAL SETUP OF SEALING COMPOUND WILL OCCUR IN APPROXIMATELY 30 MI...

Страница 20: ...oduct is intended for use indoors only POLLUTION DEGREE INSTALLATION CATEGORY ALTITUDE HUMIDITY ELECTRICAL SUPPLY VOLTAGE TOLERANCE AMBIENT TEMPERATURE RANGE PRODUCT USAGE Refer to IEC 664 1 Temperatu...

Страница 21: ...any use not Do Caution coil cooling the puncturing of danger real very the of because surfaces refrigerator on and frost ice remove to scraper or blade instrument sharp any use not Do Caution Materia...

Страница 22: ...ted to 1 4 defrost Keep containers covered when possible to reduce moisture buildup Set operating temperature no colder than necessary for the items being refrigerated Safety Tips After a unit is in o...

Страница 23: ...performed by a qualified service representative Cleaning of Units Turn off the freezer Wash interior liner with a solution of two tablespoons of baking soda and a quart of warm water Then rinse and w...

Страница 24: ...is defrost unit A leak around the door gasket will allow cold air to escape This causes unit to work harder than necessary to maintain cold temperatures Re seat or replace the gasket if worn If you h...

Страница 25: ...replacement parts information and pricing please call the Customer Service Department at 1 800 438 4851 and have the unit s model serial and code numbers available This information is located on data...

Страница 26: ...ientific dealer from whom you purchased this unit or can be obtained promptly from the factory When service or replacement parts are needed we ask that you check first with your dealer If the dealer c...

Страница 27: ...sungfallen finden Sie unter www thermo com Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2002 96EC in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici WE...

Страница 28: ...ct by a party other than the manufacturer or its duly authorized service representative ii misuse including use inconsistent with written operating instructions for the product mishandling contaminati...

Страница 29: ...can be found on the dataplate attached to the equipment If available please provide the date purchased the source of purchase manufacturer or specific agent rep organization and purchase order number...

Страница 30: ...Thermo Fisher Scientific Inc 275 Aiken Road Asheville NC 28804 United States www thermofisher com 327342H01 Rev F...

Отзывы: