2
Connecting the power
WARNING: Before connecting the FastPrep
®
Instrument to an outlet. Make sure
that the voltage, frequency, and amperage are compatible with the requirements
indicated on the label on the instrument rear panel. If there are any questions,
please consult an electrician.
As a safety feature, this unit is equipped with a three-prong grounded plug that fits
a grounding-type power outlet. Consult an electrician to replace the outlet if neces-
sary. Do not defeat this safety feature by modifying the plug. Make sure the power
switch, located on the rear panel, is OFF. Connecting the power cord to the instru-
ment (power fixture is on the rear panel) and plug it into a compatible outlet.
Note: A safety latch prevents the FastPrep
®
FP120 lid from opening when the main
power switch, located on the rear panel, is OFF. Before switching power ON, make
sure that the unit is properly set up, connected to a suitable power source, and the
RUN/OFF/CHANGE switch on the front control panel is turned OFF (see Figure 1.
Below). Once the main power is ON, press COVER (SPEED/TIME/COVER switch)
to release the lid.
2.0
INSTALLATION (cont’d)
3.0
OPERATING INSTRUCTIONS
3.1
PREPARING THE FASTPREP
®
FP120 FOR OPERATING
Turn the main power switch ON. This switch is located on the rear panel of the instru-
ment. When switched O N, factory-set default values automatically program
FastPrep to operate at speed 4.0 m/sec and run time 20 seconds (to override these
values, see section 3.2 below).
3.2
CONTROL PANEL
There are two rocker switches on the instrument front panel (Figure 1). The
RUN/OFF/CHANGE switch on the right (switch B) allows the user to choose RUN,
for operation (cellular disruption) or CHANGE, to program the desired operating
parameters (speed or time). Setting the switch to the OFF (center) position opens
the lid after a run is completed. The OFF position can also be used, if necessary, to
terminate a run that is in progress.
Note: Run settings cannot be changed once the run is in progress.
A
B
DISPLAY
Figure 1. FastPrep
®
FP120 Control Panel