Thermo Savant FastPrep FP110 Скачать руководство пользователя страница 4

2

Connecting the power

WARNING: Before connecting the FastPrep

®

Instrument to an outlet. Make sure

that the voltage, frequency, and amperage are compatible with the requirements
indicated on the label on the instrument rear panel. If there are any questions,
please consult an electrician.

As a safety feature, this unit is equipped with a three-prong grounded plug that fits
a grounding-type power outlet. Consult an electrician to replace the outlet if neces-
sary. Do not defeat this safety feature by modifying the plug. Make sure the power
switch, located on the rear panel, is OFF. Connecting the power cord to the instru-
ment (power fixture is on the rear panel) and plug it into a compatible outlet.

Note:  A safety latch prevents the FastPrep

®

FP120 lid from opening when the main

power switch, located on the rear panel, is OFF. Before switching power ON, make
sure that the unit is properly set up, connected to a suitable power source, and the
RUN/OFF/CHANGE switch on the front control panel is turned OFF (see Figure 1.
Below). Once the main power is ON, press COVER (SPEED/TIME/COVER switch)
to release the lid.

2.0 

INSTALLATION (cont’d)

3.0 

OPERATING INSTRUCTIONS

3.1

PREPARING THE FASTPREP

®

FP120 FOR OPERATING

Turn the main power switch ON. This switch is located on the rear panel of the instru-
ment.  When  switched  O N,  factory-set  default  values  automatically  program
FastPrep to operate at speed 4.0 m/sec and run time 20 seconds (to override these
values, see section 3.2 below).

3.2

CONTROL PANEL

There are two rocker switches on the instrument front panel (Figure 1).  The
RUN/OFF/CHANGE switch on the right (switch B) allows the user to choose RUN,
for operation (cellular disruption) or CHANGE, to program the desired operating
parameters (speed or time).  Setting the switch to the OFF (center) position opens
the lid after a run is completed.  The OFF position can also be used, if necessary, to
terminate a run that is in progress.

Note: Run settings cannot be changed once the run is in progress.

A

B

DISPLAY

Figure 1. FastPrep

®

FP120 Control Panel

Содержание FastPrep FP110

Страница 1: ...Inc North America 2251 Rutherford Road Carlsbad CA 92008 Phone 800 424 6101 Fax 760 918 9313 Qbiogene Inc Europe Parc d Innovation BP72 67042 Illkirch Cedex France Phone 33 0 388 675 425 Fax 33 0 388...

Страница 2: ...tting Target Time 4 3 3 Example of Manual Run 5 3 4 Example of Auto Run 6 3 5 Drying Rate 6 3 6 Other Topics 7 Section Section Title Specifications Page 4 0 Specifications 8 Section Section Title Acce...

Страница 3: ...675 425 Europe or visit the FastPrep Users Resource Group at http www qbiogene com 2 0 INSTALLATION Unpacking Carefully remove the FastPrep Instrument and accessories from the shipping carton Compare...

Страница 4: ...ble power source and the RUN OFF CHANGE switch on the front control panel is turned OFF see Figure 1 Below Once the main power is ON press COVER SPEED TIME COVER switch to release the lid 2 0 INSTALLA...

Страница 5: ...it returns to the OFF position Note While setting run speed the display counts forward in increments of 0 5 m sec within the operating range of 4 0 6 5 m sec B Select desired time of run The units di...

Страница 6: ...ke sure that their screw caps are tight Load sample tubes into the sample holder Whenever possible position the tubes symmetrically The FastPrep Instrument is designed to be used exclusively with the...

Страница 7: ...switch A SPEED TIME COVER to SPEED To display time remaining in the run set switch A SPEED TIME COVER to TIME Note When SPEED is selected on switch A AA is displayed during motor acceleration When th...

Страница 8: ...peed Construction Anodized aluminum sample holder Polypropylene sample retention spoke plate Chemical resistant painted surfaces Dimensions 11 in x 15 75 in x 12 in W x D x H 28 0 cm x 40 0 cm x 30 5...

Страница 9: ...of shipping the unit are not covered under warranty The warranty obliges you to follow the precautions in this manual When returning apparatus that may contain hazardous materials you must pack and l...

Страница 10: ...PENDIX 1 MAINTENANCE CLEANING The cell disruption process during a FastPrep Instrument run is caused by the collision of matrix and sample within the FastPrep Instrument sample tube The rate of collis...

Страница 11: ...interlock override handle is visible underneath situated toward the front The override mechanism is a flat metal handle with release cord wound around attached to the undersurface by two Phillips hea...

Страница 12: ...104 3003 00 Rev B 2000 Thermo Savant...

Отзывы: