![Thermo Neslab RTE 140 Скачать руководство пользователя страница 33](http://html1.mh-extra.com/html/thermo-neslab/rte-140/rte-140_instruction-and-operation-manual_1104437033.webp)
- 32 -
Remote Programming
Remote Programming
Remote Programming
Remote Programming
Remote Programming
The bath can be programmed and controlled by a remote computer. Enter
step setpoints using the SYn command, time periods using the SXn com-
mand, number of loops using the SKxxxx command and number of periods
in the program using the Ssn command. Start the program using SG1 and
suspend the program using SG0. When the program is finished it will stop
automatically. If needed, SR will end the program at any time.
Here is a four setpoint, two loop example:
SV1[CR]
Select program number one
SY1,25.00[CR]
First setpoint = 25.00 degrees
SX1,20.0[CR]
First period = 20.0 minutes
SY2,40.00[CR]
Second setpoint = 40.00 degrees
SX2,60.0[CR]
Second period = 60.0 minutes
SY3,40.00[CR]
Third setpoint = 40.00 degrees
SX3,45.0[CR]
Third period = 45.0 minutes
SY4,25.00[CR]
Fourth setpoint = 25.00 degrees
Ss3[CR]
Program has three periods
SK2[CR]
Execute the program three times total
SG1[CR]
Begin executing program
SG0[CR]
Temporarily suspend the program
SR[CR]
Stop the program before it is finished