1.3 Installing the Unit
a. Choosing the Location
Units must be installed against a wall or similar
structure. Maintain a three inch clearance behind
the incubator for electrical, water and gas hook-
ups.
• Locate the unit on a firm level surface capable of support-
ing the unit weight of approximately 500 lbs.
• Locate the unit away from doors and windows and heating
and air conditioning ducts.
This incubator weighs approximately 500 lbs.
Have sufficient personnel available when moving.
b. Preliminary Cleaning and Disinfecting
Remove the shipping tape securing the screws to the side-
wall of the chamber. Disinfect all interior surfaces with a gener-
al-use laboratory disinfectant. Rinse thoroughly with sterile dis-
tilled water, then 70% ethanol. Dry with a sterile cloth as need-
ed.
Disinfect the shelf channels and shelves, then rinse with
distilled water before installing.
Before using any cleaning or decontamination
method except those recommended by the manu-
facturer, users should check with the manufactur-
er that the proposed method will not damage the
equipment.
Accidental spills of hazardous materials on or
inside this unit are the responsibility of the user.
c. Installing the Shelves
The shelves may be
installed at any level in
the incubator. Install a
shelf channel on each
side. With the tabs point-
ing up, attach the channel
by locating the rivet into a
slotted hole, far end first.
Pull the channel toward
the front and slide the
front rivet on the channel into the slotted hole and press down.
Make sure that the channels are opposite each other so that the
installed shelf will be level.
d. Leveling the Unit
Place a bubble-type level on a shelf inside the incubator.
Adjust the feet as needed; counterclockwise to lengthen or
clockwise to shorten. Level the unit front-to-back and left-to-
right.
e. Attaching the Drain Lines
NOTE:
Drain connections on the back of the unit are shipped
capped. If humidity is not required, the caps should be left on
the drains.
If manual or gravity feed method is used for humidity in
the cabinet, connecting the drains is optional.
If an in-house distilled water supply is used to fill the
reservoir, it is recommended that the drains be connected to a
floor drain.
Model 3950/3951 ____________________________________________________________________Installation and Start-Up
1 - 3
Figure 1-5
Figure 1-6
Содержание 3950
Страница 14: ...Model 3950 3951 _____________________________________________________________________________Calibration 2 2 ...
Страница 18: ...Model 3950 3951 _____________________________________________________________________________Configuration 3 4 ...
Страница 19: ...Model 3950 3951 _____________________________________________________________________________Configuration 3 5 ...
Страница 30: ...Model 3950 3951 ______________________________________________________________________________Spare Parts 8 2 ...
Страница 31: ...Model 3950 3951 ______________________________________________________________________________Spare Parts 8 3 ...
Страница 32: ...Model 3950 3951 ______________________________________________________________________________Spare Parts 8 4 ...
Страница 33: ...Model 3950 3951 ______________________________________________________________________________Spare Parts 8 5 ...
Страница 34: ...Model 3950 3951 ______________________________________________________________________________Spare Parts 8 6 ...
Страница 35: ...Model 3950 3951 ______________________________________________________________________________Spare Parts 8 7 ...
Страница 36: ...Model 3950 3951 ______________________________________________________________________________Spare Parts 8 8 ...
Страница 37: ...Model 3950 3951 ______________________________________________________________________________Spare Parts 8 9 ...
Страница 38: ...Model 3950 3951 ______________________________________________________________________________Spare Parts 8 10 ...
Страница 43: ...Thermo Forma Millcreek Road P O Box 649 Marietta Ohio 45750 U S A Telephone 740 373 4763 Telefax 740 373 4189 ...
Страница 46: ...Thermo Forma Millcreek Road P O Box 649 Marietta Ohio 45750 U S A Telephone 740 373 4763 Telefax 740 373 4189 ...