Section 4 - Alarms
4.1 Alarms
The Model 3950 Series incubator is equipped with a sys-
tem which notifies the user of an alarm condition inside the
incubator. All alarms are displayed in the control panel message
center. The following table contains information on all possible
systems alarms.
To avoid alarms going off in day-to-day use, some alarms
are equipped with a time delay feature. For this to function cor-
rectly, the alarm condition must exist for the specified length of
time before the message center will display the alarm. This
allows for interruptions, such as door openings, to occur with-
out the incubator going into a continuous state of alarm.
When an alarm condition exists, the Silence key can be
pressed to temporarily mute the audible alarm. The message
center will continue to show the alarm condition. If the alarm
condition is not corrected within a specified length of time, the
alarm will sound again or “ringback” to remind the user.
When multiple alarm conditions occur, active messages are
displayed in the display center one at a time, updating at 5-sec-
ond intervals. Pressing Silence during multiple alarms causes
all active alarms to be muted and to ringback in 15 minutes.
The temperature alarms are disabled when the Temp set
point is 10°C. The CO
2
alarms are disabled when the CO
2
set
point is 0.0%.
Model 3950/3951 _________________________________________________________________________________Alarms
4 - 1
4.2 Sensor Fault Alarms
The microprocessor in 3950 Series incubators continually
scans all available sensors to ensure that they are operating
properly. Should an error be detected, the incubator will sound
an alarm and display the appropriate message. If such an alarm
occurs, contact your local distributor or the Thermo Forma
Service department at 740-373-4763 or 1-888-213-1790
(U.S.A. and Canada) or fax 740-373-4189.
Description
Message Code
Delay
Ringback
Relay
No alarm condition exists
SYSTEM OK
-----
-----
-----
Temp > Otemp Set point
SYS IN OTEMP
0 min.
15 min.
Yes
Air Temp Sensor Fault (See Section 4.2)
TMP SENS ERR
0 min.
15 min.
No
Temperature Controller Failure (See Section 4.2)
TMP CTRL ERR
0 min.
15 min.
YES
CO
2
Sensor Fault (See Section 4.2)
CO2 SENS ERR
0 min.
15 min.
No
Water Level in Reservoir is Low
ADD WATER
0 min.
15 min.
No
Inner Door is Open
DOOR OPEN
15 min.
15 min.
No
CO
2
> CO2 High Tracking Alarm
CO2 IS HIGH
15 min.
15 min.
Programmable
CO
2
< CO2 Low Tracking Alarm
CO2 IS LOW
15 min.
15 min.
Programmable
TEMP < TEMP Low Tracking Alarm
TEMP IS LOW
15 min.
15 min.
Programmable
Tank 1 is low, switch to Tank 2 (Gas Guard only)
TANK1 LOW
0 min.
----
No
Tank 2 is low, switch to Tank 1 (Gas Guard only)
TANK2 LOW
0 min.
----
No
Both tanks are low (Gas Guard only)
TANK 1 and 2 LOW
0 min.
15 min.
No
- All alarm delays and ringback times are + 30 seconds -
Содержание 3950
Страница 14: ...Model 3950 3951 _____________________________________________________________________________Calibration 2 2 ...
Страница 18: ...Model 3950 3951 _____________________________________________________________________________Configuration 3 4 ...
Страница 19: ...Model 3950 3951 _____________________________________________________________________________Configuration 3 5 ...
Страница 30: ...Model 3950 3951 ______________________________________________________________________________Spare Parts 8 2 ...
Страница 31: ...Model 3950 3951 ______________________________________________________________________________Spare Parts 8 3 ...
Страница 32: ...Model 3950 3951 ______________________________________________________________________________Spare Parts 8 4 ...
Страница 33: ...Model 3950 3951 ______________________________________________________________________________Spare Parts 8 5 ...
Страница 34: ...Model 3950 3951 ______________________________________________________________________________Spare Parts 8 6 ...
Страница 35: ...Model 3950 3951 ______________________________________________________________________________Spare Parts 8 7 ...
Страница 36: ...Model 3950 3951 ______________________________________________________________________________Spare Parts 8 8 ...
Страница 37: ...Model 3950 3951 ______________________________________________________________________________Spare Parts 8 9 ...
Страница 38: ...Model 3950 3951 ______________________________________________________________________________Spare Parts 8 10 ...
Страница 43: ...Thermo Forma Millcreek Road P O Box 649 Marietta Ohio 45750 U S A Telephone 740 373 4763 Telefax 740 373 4189 ...
Страница 46: ...Thermo Forma Millcreek Road P O Box 649 Marietta Ohio 45750 U S A Telephone 740 373 4763 Telefax 740 373 4189 ...