![Thermo Finnigan LCQ Deca XP Plus Скачать руководство пользователя страница 148](http://html.mh-extra.com/html/thermo-finnigan/lcq-deca-xp-plus/lcq-deca-xp-plus_hardware-manual_1104316148.webp)
Service Maintenance
Replacing PCBs and Assemblies in the Tower _____________________________________
5-42
___________________ LCQ Deca XP Plus Hardware Manual ________________
j.
Go to step 7.
7. Return the embedded computer to its original position as follows:
a.
Lift the embedded computer up and away from the MS detector
chassis. Lift up the embedded computer a sufficient distance to
access the bottom of the embedded computer.
b. Reconnect to the bottom of the embedded computer the cable that
comes from the switching power supplies.
c.
Reconnect the fan power cable to the embedded computer fan. Make
sure that the plug on the end of the cable contours the fan (that is, the
concave side of the plug is against the fan).
d. Reposition the embedded computer in its original position in front of
the tower.
e.
Tighten the six fasteners that secure the embedded computer to the
vacuum manifold, base plate, and chassis.
f.
Reconnect the cables that connect to the top of the embedded
computer. See Figure 5-14 on page 5-44 and Figure 5-15 on
page 5-47.
8. Reinstall the top cover of the MS detector as described in the topic
Reinstalling the Top Cover of the MS Detector on page 5-16.
9. Close the front doors of the MS detector.
10. Restart the system as described in the topic Starting Up the System after
a Complete Shutdown in the System Shutdown, Startup, and Reset
chapter.
11. Run the Tune Plus diagnostics to verify that the system is operational.
Replacing PCBs in the Embedded Computer
The Ethernet PCB (P/N 97033-60280), Acquisition DSP PCB
(P/N 97000-61260), Control DSP PCB (P/N 97000-61270), Waveform DDS
PCB (P/N 97000-61430), and CPU PCB (P/N 97144-60260) reside in the
embedded computer. See Figure 5-14.
To replace a PCB in the embedded computer, proceed as follows:
1. Place the electronics service switch in the Service Position (or shut down
and vent the XP Plus system as described in the topic Shutting Down the
System Completely in the System Shutdown, Startup, and Reset
chapter).