Check inner body operation
Hold electrode by outer body and pull up on electrode cable.
Internal filling solution will flow under membrane and restore
electrode response
Allow solutions to come to same temperature
before measurement
Prepare fresh standards
Use recommended CO
2
buffer, Orion No. 950210
Used CO
2
buffer!
Hold electrode by outer body and pull up on electrode
for extended period cable. Internal filling solution will flow under
membrane and restore electrode response.
Replace membrane
Check inner body operation
Plot millivolts on the linear axis. On the log axis, be sure
concentration numbers within each decade are increasing with
increasing concentration
Be sure to note sign of millivolt value correctly
Prepare fresh standards
Apply correct conversion factor:
10
-3
M = 44 ppm as CO
2
= 100 ppm as CaCO
3
Add same proportion to standards and samples