Symptom Possible
Causes
Drift (readings
Defective inner body
changing slowly
in one direction)
Electrode exposed to air
for extended period
Samples and standards
at different temperaures
Low slope or
Standards contaminated or
No slope
incorrectly made
CO
2
buffer not used
Standard used as CO
2
buffer
Electrode exposed to air
Membrane failure (wet,
perforation, discoloration)
Defective inner body
“Wrong Answer”
Incorrect scaling of
(But calibration
semilog paper
curve is OK)
Incorrect sign
Incorrect standards
Wrong units used
CO
2
buffer added to
standards and not samples