![Thermo Electron iEMS Скачать руководство пользователя страница 21](http://html.mh-extra.com/html/thermo-electron/iems/iems_user-manual_1104277021.webp)
Chapter 6: Operation
iEMS
Incubator/Shaker User Manual
Rev. 3.0, Cat. no. 1506200A
21
Program the incubation time by dialing the number keys: first hours; then
minutes, and finally seconds.
Note:
Check where the cursor (underline) is located before you enter a value.
The value will be entered at the location of the cursor.
For example, for an incubation time of 1 hour 30 minutes:
Press
01 30 00
↵
Note:
If you want to interrupt the programming, press
stop
to return to the
READY state. Your changes will be saved.
Program the shaking time by dialing the number keys: first seconds; then
hours, and finally minutes. It is recommended that the same shaking time is
used in all slots in one module. If you need different shaking times, use slots in
different modules.
Note:
Check where the cursor (underline) is located before you enter a value.
The value will be entered at the location of the cursor.
For example, for a shaking time of 20 seconds:
Press
00 00 20
↵
Note:
If you want to interrupt programming, press
stop
to return to the
READY state. Your changes will be saved.
Interval time is a programmable parameter that allows you to specify at what
interval the shaking is alternately switched off and on (e.g., an interval time of
00:00:10 results in 10 seconds of shaking turned on, 10 seconds of shaking
turned off, 10 seconds of shaking turned on, etc.) for the duration of the
programmed shaking time.
Program the interval shaking time by dialing the number keys: first minutes;
then seconds, and finally hours. It is recommended that the same interval time
is used in all slots in one module. If you need different interval times, use slots
in different modules.
Note:
Check where the cursor (underline) is located before you enter a value.
The value will be entered at the location of the cursor.
For example, for an interval shaking time of zero:
Press
00 00
↵
(Enter)
INC TIME
01:30:00
SHAKE TIME
00:00:00
SHAKE TIME
00:00:20
INT TIME
00:00:00
SHAKE TIME
00:00:20
INT TIME
00:00:00
Содержание iEMS
Страница 1: ...iEMS Incubator Shaker Rev 3 0 Cat no 1506200A iEMS Incubator Shaker User Manual ...
Страница 59: ...iEMS Incubator Shaker User Manual Rev 3 0 Cat no 1506200A 59 ...
Страница 60: ...iEMS Incubator Shaker User Manual Rev 3 0 Cat no 1506200A 60 ...