![Thermo Electron iEMS Скачать руководство пользователя страница 20](http://html.mh-extra.com/html/thermo-electron/iems/iems_user-manual_1104277020.webp)
Chapter 6: Operation
iEMS
Incubator/Shaker User Manual
Rev. 3.0, Cat. no. 1506200A
20
6.2 Programming
protocols
Use the keyboard to enter processing parameters and to start thus programmed
protocols. You can program a single protocol for all slots or an individual
protocol for each slot. The parameters are stored in the instrument memory
until you enter new parameters. You can change the parameters of one slot
without affecting the parameters of the other slots.
In one module, all slots have the same incubation temperature and they shake
at the same speed. You should preheat the slots to the programmed target
temperature before you start any protocol (see Section 6.5 Preheating and
heating).
Operation Display
Switch the power ON.
Note:
If the power is already on, any shaking or heating must be turned off in
the slot to be able to program a protocol.
To program the incubation and shaking parameters, press the appropriate slot
number (
1
−
9
). For example,
Press
1
↵
(Enter)
If you want to program the same parameters to all slots:
Press
0
↵
Program the new temperature value by dialing the number keys, or press
↵
(Enter) to accept the existing one.
Note:
Check where the cursor (underline) is located before you enter a value.
The value will be entered at the location of the cursor.
For example,
Press
30
↵
Note:
If you want to interrupt the programming, press
stop
to return to the
READY state. Your changes will be saved.
READY
09:25:15
PLACE 1
09:25:15
TEMPERATURE 37.0ºC
PLACES 1-9
TEMPERATURE 37.0ºC
TEMPERATURE 30.0ºC
INC TIME
00:00:00
Содержание iEMS
Страница 1: ...iEMS Incubator Shaker Rev 3 0 Cat no 1506200A iEMS Incubator Shaker User Manual ...
Страница 59: ...iEMS Incubator Shaker User Manual Rev 3 0 Cat no 1506200A 59 ...
Страница 60: ...iEMS Incubator Shaker User Manual Rev 3 0 Cat no 1506200A 60 ...
Страница 61: ...iEMS Incubator Shaker User Manual Rev 3 0 Cat no 1506200A 61 ...