For your safety, check that mains wiring is effectively grounded.
Thermo declines all responsibility for any damages due to non-grounding of the machine.
Remember that in order to respect the electrical safety standards related to protection against indirect
contact, the supply of power to the instrument must be via a power socket fitted with a protection device
ensuring automatic cut-off in the case of an insulation fault. A supply fitted with a circuit breaker of the
correct rating complies with this requirement.
3.5. Lid Opening and Rotor Checking
Ensure that the centrifuge has been switched ON. Pull the handle on the centrifuge right side towards the
front of the unit : the lid is automatically unlocked and opens.
In the case of a mains power outage, opening of the lid is prevented by the lid lock safety device. It is
recommended to wait for the mains power to be switched back on so that this safety device enables the lid
to be unlocked (refer to 3.6 for manual lid opening).
Carefully clean the inside of the centrifugation chamber removing any packing residues.
In fact, due to air turbulence, solid particles accidentally left in the centrifugation chamber create
excessive wear of the chamber itself and of the outer rotating equipment surface.
To install the rotor:
Carefully lower the rotor onto the drive shaft. Press down on the rotor until a click is heard. Try to lift the
rotor. When correctly placed, it will not move, being automatically locked onto the drive.
3.4. Mains Supply
Check mains and frequency: they must correspond to the values shown on the instrument identification
label.
C3i-CR3i
Installation
8
Cat. No.
Content
Voltage
Frequency
Type
11175770
C3i
230V ± 10%
50/60 Hz
10 A single phase + ground
11175774 CR3i
11175771
C3i
120V +5% -10%
60 Hz
16 A single phase + ground
11175775
CR3i
It is not necessary to orientate the rotor relative to the drive shaft in order to achieve locking. The AUTO-
LOCK rotor mounting system allows rotors to be placed in any orientation.
1)
“clic”
2)