Flow chart of disinfection routine:
Status display on CO
2
display:
Heat:
The chamber is heated to 180°C; the disinfection temperature can be read off the temperature
display.
Disinfect:
When the disinfection temperature of 180°C is reached, a disinfection phase with a duration of
approx. 3 hours is started. If the door is opened during the period and the temperature drops
below 180°C, the disinfection routine restarts automatically.
Cool:
On completion of the disinfection phase, the unit cools down to the originally set nominal
temperature.
The yellow indicator lamp "des" as soon as the unit reaches the set operating temperature
(e.g. 37.0°C) is reached. The sensors are automatically calibrated (auto-start) and the unit
is readjusted to the set gas / rel. humidity setpoints. The disinfection routine is completed.
If a temperature of 200°C is exceeded during the disinfection routine, the latter is cancelled
and all poles of the heating system are disconnected. Error message 501 is displayed.
Cancelling the disinfection routine:
Instruction
Entry / Keys / Remarks
Display / Remarks / Status
Cancel disinfection
des
start/
stop
The CO
2
display indicates the cool-
ing phase of the disinfection routine.
Press for approx. 1 second
6. OPERATE
22/43
50 079 041
Start
Time
37°C
180°C
Temp.
End
Heating phase
Disinfection
phase
Cooling phase