When replacing the hose, do not touch the temperature sensors. If
these sensors are touched in any way, the unit must be recalibrated.
The Mistral-Air
®
Plus warming unit is fitted with an air filter; however
airborne contamination shall be taken into consideration when using
the warming system.
In order to prevent the patient from becoming hypothermic due to
unanticipated loss of power, connect the unit to the emergency
mains power.
When the unit is switched off, use of good thermal conductivity
materials on the patient could decrease the patient temperature.
The heating device does not contain an alarm system with an
interruption of power supply/supply mains alarm condition. This
means that in case of a power failure, there will be no alarm.
The unit is not equipped with an isolating switch. Temporary
interruption of the supply mains will render the device in Stand-by
Mode.
The Mistral-Air
®
Plus warming unit is certified according to IEC
60601-1-2
:2014
for electromagnetic interference. However, if
electromagnetic interference with nearby devices is experienced, the
user is encouraged to correct the interference by one or more of the
following measures:
•Isolate the offending device.
•Reorient or relocate the Mistral-Air
®
Plus warming unit.
•Increase the distance between the interfering device and the
Mistral-Air
®
Plus warming unit, use another electrical outlet.
•If assistance is required, please contact your local dealer.
Use of Mistral-Air® Plus warming unit adjacent to or stacked with
other equipment should be avoided because it could result in
improper operation. If such use is necessary, the Mistral-Air® Plus
warming unit and the other equipment should be observed to verify
that they are operating normally.
Do not use the Mistral-Air
®
Plus warming unit with any forced air
disposables other than Mistral-Air
®
blankets. Thermal injury may
result.
Do not place the unit
in
or on the bed with the patient.
A physician order is required for setting temperature and for
continued use.