![The37Company Mistral-Air MA1100-EU Скачать руководство пользователя страница 21](http://html.mh-extra.com/html/the37company/mistral-air-ma1100-eu/mistral-air-ma1100-eu_user-manual_1101095021.webp)
INT-R295-EN-10-11-17
18
By pressing the “-” of the Temperature Selecting Button, the temperature
setting decreases. This is indicated by a green LED:
After selecting the desired temperature, the LED below the Temperature
Indicator will flash green. After reaching the set temperature (+ 2 ˚C) the
green flashing LED will light up permanently.
8.5
Stop Warming
a
b
c
a. Press the Stand-by button
b. Disconnect the hose from the blanket
c. If desired, leave the blanket on/under the patient
9
Safety Systems and Alarms
The Mistral-Air
®
Plus
warming unit is equipped with visual and audible
safety systems to protect against over temperature and under temperature
conditions as well as to indicate that filter change is required.
In case of temperature alarm, check for free airflow. Ensure
blanket is not folded and do not place tools/equipment on the
blanket which could result in a blocked air flow. Be sure the air
inlet is free. If the unit continues to alarm, take the unit out of use
and contact the hospital service department or the local supplier.