5.1.2.3 Cable entry and mounting
The E-Scan console is set up to be mounted to a wall or other mounting device by the bracket
located on the bottom of the console. This also serves as a cable and power entry plate. It is
possible to change out the console without removing any wiring from the connectors. This can
be done by removing the cable connectors from the internal interface board along with the
screws holding the console to the mounting plate. Another method may be to utilize the stainless
steel connector option offered for the sensing head.
5.1.2.4 IB (interface board) for external connection
The interface board is located on the back wall of the console. Most external connections are
made to connectors that plug into this board. The only exceptions are optional features that
require connections to optional available connectors located on the side of the console.
5.1.2.5 Power supply
The power supply is located in the top of the console and serves as a hold down clamp for the
analog output modules. The power supply supplies power to both the I/O card and also the IB.
The voltages supplied are +5 vdc and +/- 15 vdc.
5.1.2.6 I/O card
The I/O card handles most of the signal inputs and outputs from the instrument. Two ribbon
cables route these signals from the IB to the I/O card and power is supplied through a power
cable from the power supply.