Unpack

Connect the instrument

Important test system safety information

This product is sold as a stand-alone instrument that may 

become part of a system that could contain hazardous 

voltages and energy sources. It is the responsibility of the 

test system designer, integrator, installer, maintenance 

personnel, and service personnel to make sure the system is 

safe during use and is operating properly.

You must also realize that in many test systems a single 

fault, such as a software error, may output hazardous signal 

levels even when the system indicates that there is no 

hazard present.

It is important that you consider the following factors in your 

system design and use:

The international safety standard IEC 61010-1 defines

voltages as hazardous if they exceed 30 V

RMS

 and

42.4 V

PEAK

 or 60 V dc for equipment rated for dry

locations. Keithley Instruments products are only rated for

dry locations.

Read and comply with the specifications of all instruments

in the system. The overall allowed signal levels may be

constrained by the lowest rated instrument in the system.

For example, if you are using a 500 V power supply with

a 300 V dc rated switch, the maximum allowed voltage in

the system is 300 V dc.

• Cover the device under test (DUT) to protect the operator

from flying debris in the event of a system or DUT failure.

Make sure any test fixture connected to the system

protects the operator from contact with hazardous

voltages, hot surfaces, and sharp objects. Use

shields, barriers, insulation, and safety interlocks to

accomplish this.

• Double-insulate all electrical connections that an operator

can touch. Double insulation ensures the operator is

still protected even if one insulation layer fails. Refer to

IEC 61010-1 for specific requirements.

• Make sure all connections are behind a locked cabinet

door or other barrier. This protects the system operator

from accidentally removing a connection by hand and

exposing hazardous voltages. Use high-reliability fail-safe

interlock switches to disconnect power sources when a

test fixture cover is opened.

• Where possible, use automatic handlers so that operators

are not required to access the DUT or other potentially

hazardous areas.

• Provide training to all users of the system so that they

understand all potential hazards and know how to protect

themselves from injury.

• In many systems, during power up, the outputs may be in

an unknown state until they are properly initialized. Make

sure the design can tolerate this situation without causing

operator injury or hardware damage.

Содержание Keithley System SourceMeter 2601B-PULSE

Страница 1: ...Model 2601B PULSE System SourceMeter Instrument Quick Start Guide ...

Страница 2: ...r high transient overvoltages often associated with local AC mains connections Certain Keithley measuring instruments may be connected to mains These instruments will be marked as category II or higher Unless explicitly allowed in the specifications operating manual and instrument labels do not connect any instrument to mains Exercise extreme caution when a shock hazard is present Lethalvoltage may...

Страница 3: ...ad the associated information very carefully before performing the indicated procedure The CAUTION heading in the user documentation explains hazards that coulddamage the instrument Such damage may invalidate the warranty The CAUTION heading with the symbol in the user documentation explains hazards that could result in moderate or minor injury or damage the instrument Always read the associated i...

Страница 4: ...maximum at 6 V dc 1 A maximum at 40 V dc Pulser feature output region 4 Region maximums 10 A at 20 V Maximum pulse width 1 8 ms Maximum duty cycle 1 Measurement input electrical ratings Measurement category O Voltage 40 V dc maximum HI to LO Current 3 A at 6 V dc 1 A maximum at 40 V dc Impedance Variable CAUTION Carefully consider and configure the appropriate output off state source levels and co...

Страница 5: ...hooting and programming commands of the instrument Information on accessories Software for the 2601B PULSE is also available for download from the Keithley web page at tek com keithley You can search for the specific software you need Available software includes Keithley KickStart Instrument Control Software Lets you start making measurements in minutes without complex instrument programming Free ...

Страница 6: ... You receive the 2601B PULSE with these accessories and documents 1 Interlock DB 25 male connector kit hardware 2 Power line cord 3 Safety interlock mating connector 4 Two RJ 45 LAN crossover cables 5 2601B P INT Rear Panel Interlock and Cable Connector Box not shown 6 Safety supplement not shown 7 Safety precautions not shown 8 Software and documentation downloads document not shown 9 Model 2601B...

Страница 7: ... allowed voltage in the system is 300 V dc Cover the device under test DUT to protect the operator from flying debris in the event of a system or DUT failure Make sure any test fixture connected to the system protects the operator from contact with hazardous voltages hot surfaces and sharp objects Use shields barriers insulation and safety interlocks to accomplish this Double insulate all electric...

Страница 8: ...ottom of the instrument Position the instrument so that it is easy to reach any disconnecting devices such as the power cord and the power switch Install the 2601B P INT The 2601B PULSE is shipped with the 2601B P INT Rear Panel Interlock and Cable Connector Box The 2601B P INT provides connections for an optional hardware safety interlock and simplifies test connections to the 2601B PULSE To inst...

Страница 9: ...s firmly onto the terminal strips on the rear panel of the 2601B PULSE 6 Use the two 6 32 in screws provided with the 2601B P INT to secure it to the rear panel of the 2601B PULSE Wiring the interlock You can use the 2601B P INT and the 2601B PULSE digital I O port or external wiring to make interlock connections to an external device When properly connected the output of the instrument will turn ...

Страница 10: ...event of a failure not using a properly grounded protective earth and grounded outlet may result in personal injury or death due to electric shock Do not replace detachable mains supply cords with inadequately rated cords Failure to use properly rated cords may result in personal injury or death due to electric shock CAUTION Operating the instrument on an incorrect line voltage may cause damage to...

Страница 11: ...ent is turned on you should see A series of dots All segments of the display light A brief display showing the instrument model 2601B PULSE Line frequency detection and other startup checks The entire power up process takes approximately 30 seconds to complete When initialization is complete you see the default display screen shown below SrcA 000 000mV LimA 1 00000 A AUTO ...

Страница 12: ...ol CURSOR keys Step 1 Set source function range and level 1 Press the SRC key You will see a blinking character in the SrcA value field Confirm that mV is displayed if not press the SRC key again 2 While that character is still blinking press the up or down RANGE keys until 40 V is displayed AUTO The main display screen is refreshed SrcA 00 0000 V LimA 1 00000 A AUTO 3 Press the CURSOR keys to mov...

Страница 13: ... 10 0000 mA SrcA 20 0000 V LimA 10 0000mA AUTO Source limit set to 10 0000mA Step 3 Set measurement function and range 1 Press the MEAS key until the V voltage measurement function is selected In the following figure the measurement function has been set to V SrcA 20 0000 V LimA 10 0000mA AUTO Measurement function set to voltage V 2 Press the AUTO key as many times as needed to select the AUTO ran...

Страница 14: ...the 20 V source value Step 6 Turn the output off When you finish making measurements turn the output off by pressing the OUTPUT ON OFF control The output indicator LED turns off NOTE These steps confirm basic functionality of your instrument Turn the instrument power OFF now To use the pulser feature you must have remote communications set up Refer to the Model 2601B PULSE Reference Manual documen...

Страница 15: ...n the instrument on If an error message is displayed press the EXIT LOCAL key The 2601B PULSE will return to the default display screen For detailed information about error messages see Errors and status messages in the Model 2601B PULSE Reference Manual Next steps For more information refer to the Keithley Instruments website tek com keithley for support and additional information about the instr...

Страница 16: ...S and foreign patents issued and pending Information in this publication supersedes that in all previously published material Specfication and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tektronix Inc All other trade names referenced are the service marks trademarks or registered trademarks of their respective companies 2601B PULSE 903 01 Rev A April 2020 P2601B...

Отзывы: