background image

Specifications

Measurement Category: Category III, 600 VAC

and 600 VDC

Operating Conditions:

Temperature:

°

C to 30 

°

C (32 

°

F to 86 

°

F),

0% to 80% relative humidity

30 

°

C to 40 

°

C (86 

°

F to 104 

°

F),

0% to 75% relative humidity

40 

°

C to 50 

°

C (104 

°

F to 122 

°

F),

0% to 45% relative humidity

Altitude: 2000 m (6500') maximum
Indoor use only

Storage Conditions: –20 

°

C to 60 

°

C (–4 

°

F to 140 

°

F),

0% to 80% relative humidity

Pollution Degree: 2

Measurement Categories

These definitions were derived from the international
safety standard for insulation coordination as it applies to
measurement, control, and laboratory equipment. These
measurement categories are explained in more detail by
the International Electrotechnical Commission; refer to
either of their publications: IEC 61010-1 or IEC 60664.

Measurement Category I

Signal level. Electronic and telecommunication equipment,
or parts thereof. Some examples include transient-protected
electronic circuits inside photocopiers and modems.

Measurement Category II

Local level. Appliances, portable equipment, and the
circuits they are plugged into. Some examples include
light fixtures, televisions, and long branch circuits.

Measurement Category III

Distribution level. Permanently installed machines and the
circuits they are hard-wired to. Some examples include
conveyor systems and the main circuit breaker panels of a
building’s electrical system.

Measurement Category IV

Primary supply level. Overhead lines and other cable
systems. Some examples include cables, meters,
transformers, and other exterior equipment owned by the
power utility.

Cleaning

Periodically wipe the leads with a damp cloth and mild
detergent; do not use abrasives or solvents.

Lifetime Limited Warranty

Greenlee Textron warrants to the original purchaser of these goods for
use that these products will be free from defects in workmanship and
material for their useful life, excepting normal wear and abuse. This
warranty is subject to the same terms and conditions contained in
Greenlee Textron’s standard one-year limited warranty.

For all Test Instrument repairs, contact Customer Service at
800-435-0786 and request a Return Authorization. An Approved
Authorization including shipping label and instructions will be sent.

For items not covered under warranty (such as items dropped, abused,
etc.), a repair cost quote is available upon request.

Note: Prior to returning any test instrument, please check replaceable
batteries or make sure the battery is at full charge.

Electric shock hazard:

• Do not apply more than the rated voltage between any

two input terminals, or between any input terminal and
earth ground.

• Do not contact the test lead tips or any uninsulated

portion of the accessory.

Failure to observe these warnings could result in severe
injury or death.

Electric shock hazard:

• Do not attempt to repair this unit. It contains no user-

serviceable parts.

• Do not expose the unit to extremes in temperature or

high humidity. Refer to “Specifications.”

Failure to observe these precautions may result in injury
and can damage the unit.

Important Safety Information

Отзывы: