Using the Features

All Models

• 

REC:

  Press momentarily to activate MAX/MIN/AVG recording mode. 

 

and “MAX 

MIN AVG” appear on the display. The meter beeps when new maximum or minimum 

reading is updated. The average reading is calculated over time. Press the button 

momentarily to read the MAX, MIN and AVG readings in sequence. Press the button 

for 1 second to exit MAX/MIN/AVG recording mode. APO is disabled automatically in 

this mode.

• 

SELECT:

  Press momentarily to toggle between functions, measurement modes, 

or display modes. The last used setting becomes the default when that function is 

selected again. To change the default, select a new setting. The setting will be stored 

in nonvolatile memory.

• 

HOLD:

  Press momentarily to hold the present value on the display. Press again to exit 

this mode.

• 

Intelligent Auto Power Off (APO):

  The meter automatically turns off after approxi-

mately 32 minutes of inactivity to extend battery life. Active meter function is specified 

as: selector switch is rotated or pushbuttons are pressed; readings above 8.5% of 

the range; non-OL readings for resistance, continuity, or diode function; non-zero 

readings for Hz function; phase rotation segment moving. To wake up the meter from 

Auto Power Off, press the 

SELECT

 button momentarily and release or turn the rotary 

switch 

OFF

 and then back on. Always turn the rotary switch to the 

OFF

 position when 

the meter is not in use.

CM-860, CM-960, CMH-600, and CMI-600 Only

• 

:

  Press the 

SELECT

 button for 1 second to toggle the LCD backlight. The back-

light automatically turns off after 32 seconds to extend battery life.

• 

∆:

  Press the 

∆ 

button for 1 second to toggle between zero and normal mode. In zero 

mode the value on the LCD when the button is pressed becomes the offset. As the 

input value changes, the offset is subtracted from the actual value. This modified 

value is shown on the LCD.

Note: 

∆ 

operation when DCA or DC + ACA functions are selected offsets any mag-

netic hysteresis in the jaw. If DCA or DC + ACA reading before measuring is not zero, 

press the 

∆ 

button for 1 second. The LCD will display “dc_0”. When this clears, the 

meter is ready for operation. When measuring more than 5 A, this mode is disabled. 

As an indication, the meter will beep three times.

CM-960, CMH-600, and CMI-600 Only

• 

PEAK-RMS:

  Press the 

PEAK-RMS

 button for 1 second to activate PEAK-RMS mode 

to capture peak current or voltage in duration as short as 80 ms. 

 appears on 

the display. Press the button for 1 second to exit PEAK-RMS mode. APO is disabled 

automatically in this mode.

CMH-600 and CMI-600 Only

•    and  :  If desired, press and hold the 

REC

 button while turning the meter ON 

to enable the Beeper feature. If the segments indicate a clockwise movement, the 

beeper sounds a single long beep per segment cycle. If the segments indicate a coun-

terclockwise movement, the beeper sounds three short beeps per segment cycle.

•   :  Ensure the motor is electrically isolated from its power source. Press the 

SELECT

 button momentarily to select Motor function. Connect the meter to the 

motor as described in the “Settings Table” section of this manual. From the 

perspective of looking down the shaft of the motor, spin it clockwise. If the meter 

indicates a clockwise movement, the motor leads connected to L1, L2, and L3 of 

the meter are T1, T2, and T3, respectively. If the meter indicates a counterclock-

wise movement, swap any two connects between the meter and motor. Then 

retest.

• 

:  Press the 

SELECT

 button momentarily to select Phase function. Connect the 

meter to the phases as described in the “Settings Table” section of this manual. 

If the meter indicates a clockwise movement, the phases connected to L1, L2, 

and L3 of the meter are L1, L2, and L3, respectively. If the meter indicates a 

counterclockwise movement, swap any two connects between the meter and 

phases. Then retest.

Содержание Greenlee CM-660

Страница 1: ...3 Greenlee Textron Inc 12 13 Clamp Meters INSTRUCTION MANUAL CM 660 CM 860 CM 960 CMH 600 CMI 600 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www Test...

Страница 2: ...se Rotation Peak RMS Capture Capacitance Backlight Zero Mode Other capabilities include Hold Mode to capture the present LCD value Record Minimum Maximum or Average Mode Intelligent Auto Power Off Aut...

Страница 3: ...r property damage Read and understand this material before operating or servicing this equipment Failure to understand how to safely operate this tool could result in an accident causing serious injur...

Страница 4: ...ood condition Do not use the test lead if the contrasting inner layer of insulation is visible Use this unit for the manufacturer s intended purpose only as described in this manual Any other use can...

Страница 5: ...accurate readings Failure to observe these warnings could result in severe injury or death Electric shock hazard Do not change the measurement function while the test leads are connected to a componen...

Страница 6: ...gative common or ground input terminal L1 input for motor and phase rotation 8 Refer to explanation under Using the Features 9 Jaw lever 10 Keep hand or fingers below the dashed line during operation...

Страница 7: ...cles per second 14 Ohm 15 k Kilo 103 16 AmpTip function is enabled 17 Relative zero function is enabled 18 Polarity indicator 19 VFD AC measurement with filter for accuracy on variable frequency drive...

Страница 8: ...he actual value This modified value is shown on the LCD Note operation when DCA or DC ACA functions are selected offsets any mag netic hysteresis in the jaw If DCA or DC ACA reading before measuring i...

Страница 9: ...th of the probe tip to 4 mm for measurement categories III and IV These test leads include a cap that must be in place when used in measurement category III or IV applications Failure to observe this...

Страница 10: ...rue RMS provides much more measurement versatility The Greenlee CM 660 CM 860 CM 960 CMH 600 and CMI 600 are true RMS meters The Waveforms and Crest Factors table shows some typical AC signals and the...

Страница 11: ...ncy noise on a 50 60 Hz power line Operation Electric shock hazard Contact with live circuits could result in severe injury or death Operating Procedure 1 Refer to the Settings Table Set the selector...

Страница 12: ...00 Only Capacitance F COM CM 960 and CMI 600 Only Voltage AC DC COM Current AmpTip DC A N A N A Current AmpTip AC DC A N A N A Current DC A N A N A Current AC DC A N A N A CMH 600 and CMI 600 Only Tem...

Страница 13: ...5 1 at full scale Crest factor 5 0 1 at half scale Frequency Function Sensitivity Sine RMS Range 600 V 50 V 5 00 Hz 999 9 Hz 60 0 A AmpTip 40 A 50 00 Hz 400 0 Hz 60 00 A 600 0 A 40 A 50 00 Hz 400 0 Hz...

Страница 14: ...60 00 A 1 5 0 05 A For CM 660 CM 860 and CMH 600 only For CM 960 and CMI 600 only induced error for adjacent current carrying conductors is 0 01 A A AmpTip DC Current CM 960 and CMI 600 only Range Ac...

Страница 15: ...fied 0 20 A to 5 99 A 1 8 0 15 A 2 0 0 15 A 6 00 A to 60 00 A 1 8 0 05 A 2 0 0 05 A 600 0 A 60 0 A to 600 0 A 1 8 0 5 A 2 0 0 5 A For any measurement when the conductor is not positioned on the center...

Страница 16: ...VDC Continuity Tone Threshold Between 10 and 250 Response Time Approximately 32 ms Capacitance Measurement Capacitance CM 860 CM 960 CMH 600 and CMI 600 only Range Accuracy 200 0 F 2 0 0 4 F 2500 F 2...

Страница 17: ...e 2 000 V Accuracy 1 5 0 005 V Test Current typical 0 3 mA Open Circuit Voltage typical 3 5 VDC Temperature SELECT Temperature CMH 600 and CMI 600 Range Accuracy 40 0 C 99 9 C 1 0 0 8 C 100 C 400 C 1...

Страница 18: ...e a dedicated switch position A Non contact OR B Contact precision measurement Non contact EF Detection Typical Voltage Range at 50 Hz to 60 Hz LCD Indication 20 V 10 V 36 V 55 V 23 V 83 V 110 V 59 V...

Страница 19: ...e all models and DC A mode CMH 600 and CMI 600 Total Accuracy Specified Accuracy 25 digits In an RF field of 3 V m all other modes and models not specified above Total Accuracy Specified Accuracy 20 d...

Страница 20: ...long branch circuits Measurement Category III Distribution level Permanently installed machines and the circuits they are hard wired to Some examples include conveyor systems and the main circuit bre...

Страница 21: ...placing the Batteries 1 Disconnect the unit from the circuit Turn the unit OFF 2 Remove the two screws from the battery access door and remove it 3 Replace the batteries making sure to observe the pol...

Страница 22: ...or their useful life excepting normal wear and abuse This warranty is subject to the same terms and conditions contained in Greenlee Textron Inc s standard one year limited warranty For items not cove...

Страница 23: ......

Отзывы: