CM-450

Operation

Electric shock hazard:

Contact with live circuits could result in severe injury or 
death.

1. Set the Selector switch according to the Settings Table. Momentarily press the

SELECT

 button to select mode.

2. Refer to “Typical Measurements” for specific measurement instructions.

3. Test the unit on a known functioning circuit or component.

If the unit does not function as expected on a known functioning circuit,
replace the battery.

4. Take the reading from the circuit or component to be tested.

Settings Table

To measure 

this value …

Set the Selector 

switch to this 

symbol …

This icon 

will appear 

on the 

display …

Connect 
red lead 

to …

Connect 

black 

lead 

to …

AC Current* 

(600 A max.)

 and 

A

N/A

N/A

Frequency of a 

current 

(600 A max.)

and press 

Hz

Hz

N/A

N/A

Voltage 

(600 V max.)

 and 

V

+

COM

 and 

V

Frequency of a 

voltage signal

 or 

and press 

Hz

Hz

+

COM

Resistance

+

COM

Continuity**

and press 

SELECT

Ω

 

+

COM

Diode

and press 

SELECT 

2 times

and 

V

+

COM

* AC current measurements are made using the jaw.

**  Tone indicates continuity. The threshold is between 5 Ω and 120 Ω.

www.calcert.com

[email protected]

1.800.544.2843

0

5

10

15

20

25

30

Содержание 783310127260

Страница 1: ...aparecen en este manual antes de manejar estas herramientas o darles mantenimiento Lire attentivement et bien comprendre toutes les instructions et les informations sur la s curit de ce manuel avant d...

Страница 2: ...discard this product or throw away All speci cations are nominal and may change as design improvements occur Greenlee Textron Inc shall not be liable for damages resulting from misapplication or misu...

Страница 3: ...t in severe injury or death Hazards which if not avoided COULD result in severe injury or death Hazards or unsafe practices which if not avoided MAY result in injury or property damage Read and unders...

Страница 4: ...provided by the unit Failure to observe these warnings could result in severe injury or death Electric shock hazard Do not operate with the case open Before opening the case remove the test leads or j...

Страница 5: ...Do not change the measurement function while the test leads are connected to a component or circuit Failure to observe this precaution may result in injury and can damage the unit Statement of Confor...

Страница 6: ...r 18 Relative measurement is enabled 19 M Mega 106 20 Ohms 21 Hz Hertz 22 k kilo 103 23 A Amps 24 V Volts 25 m milli 10 3 26 O L Overload indicator 27 Negative polarity indicator 28 AC measurement is...

Страница 7: ...y give higher sensitivity Reading may be zero when sensitivity is too low Electrical noise may cause unstable readings Hold Button HOLD Press momentarily to hold the present value on the display will...

Страница 8: ...RMS reading method uses internal circuitry to read the true RMS value This method is accurate within the speci ed crest factor limitations whether the input signal is a pure sine wave square wave tria...

Страница 9: ...eading from the circuit or component to be tested Settings Table To measure this value Set the Selector switch to this symbol This icon will appear on the display Connect red lead to Connect black lea...

Страница 10: ...wire in the jaw for highest accuracy Clamp Around Line Splitter Notes The Greenlee 93 30 Line Splitter is divided One section renders amps the other renders amps multiplied by 10 Close the jaw comple...

Страница 11: ...11 CM 450 Voltage Typical Measurements Resistance www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30...

Страница 12: ...12 Forward Bias Reverse Bias Typical Measurements Continuity Diode Test www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30...

Страница 13: ...equency Range 0 40 to 40 00 A 1 5 0 08 A 50 to 60 Hz 40 0 to 400 0 A 1 5 0 8 A 400 to 600 A 1 5 8 A Adjacent conductor in uence 0 05 A A Add 2 if the conductor is beyond the position error marks Multi...

Страница 14: ...ction Sensitivity Sine RMS Range 400 0 mV 350 mV 10 Hz to 2 kHz 4 000 V 1 V 10 Hz to 5 kHz 40 00 V 32 V 10 Hz to 100 kHz 400 0 V 100 V 10 Hz to 10 kHz 600 V 500 V 10 Hz to 5 kHz 400 0 A 60 A 40 Hz to...

Страница 15: ...measurement categories are explained in more detail by the International Electrotechnical Commission refer to either of their publications IEC 61010 1 or IEC 60664 Measurement Category I Signal level...

Страница 16: ...case open Before opening the case remove the test leads or jaw from the circuit and shut off the unit Failure to observe these warnings could result in severe injury or death 1 Disconnect the unit fro...

Страница 17: ...Avant d ouvrir le bo tier retirer les ls d essai ou la pince du circuit et mettre l appareil hors tension L inobservation de ces consignes pourrait entra ner des blessures graves voire mortelles 1 D b...

Отзывы: