13

CM-450

Accuracy

Refer to “Specifications” for operating conditions and temperature coefficient. 

Accuracy is specified as follows: ± (a percentage of the r a fixed amount) 
at 23 °C ± 5 °C (73.4 °F ± 9 °F), 0% to 75% relative humidity.

True RMS readings: Voltage is specified from 5% to 100% of the range, and AC 
current is specified from 10% to 100% of the range, unless otherwise specified. 
Frequency must be within the specified bandwidth for non-sinusoidal waveforms. 
Crest factors are as follows:

Crest Factor < 1.75 at full scale

Crest Factor < 3.5 at half scale

AC Current

Measurement Range

Accuracy

Frequency Range

      0.40 to 40.00 A

    ± (1.5% + 0.08 A)

50 to 60 Hz

   40.0 to 400.0 A

  ± (1.5% + 0.8 A)

   400 to 600 A

  ± (1.5% + 8 A)

Adjacent conductor influence < 0.05 A/A

Add 2% if the conductor is beyond the position error marks.

Multiply the fixed error amount times 2 below 10% of range.

AC Voltage

Measurement 

Range

Accuracy

Frequency 

Range

Input 

Impedance

   4.000 V

 ± (1.5% + 0.005 V)

50 to 500 Hz

10 MΩ // 30 pF 

nom.

   40.00 V

 ± (1.5% + 0.05 V)

   400.0 V

 ± (1.5% + 0.5 V)

   600 V

 ± (2.0% + 5 V)

DC Voltage

Measurement Range

Accuracy

Input Impedance

     400.0 mV

  ± (0.3% + 0.4 mV)

1000 MΩ*

     4.000 V

  ± (0.5% + 0.003 V)

10 MΩ

     40.00 V

  ± (0.5% + 0.03 V)

     400.0 V

  ± (0.5% + 0.3 V)

     600 V

  ± (1.0% + 4 V)

 

*  The high impedance on this range will produce significant non-zero readings 

when the test leads are disconnected from a circuit.

www.calcert.com

[email protected]

1.800.544.2843

0

5

10

15

20

25

30

Содержание 783310127260

Страница 1: ...aparecen en este manual antes de manejar estas herramientas o darles mantenimiento Lire attentivement et bien comprendre toutes les instructions et les informations sur la s curit de ce manuel avant d...

Страница 2: ...discard this product or throw away All speci cations are nominal and may change as design improvements occur Greenlee Textron Inc shall not be liable for damages resulting from misapplication or misu...

Страница 3: ...t in severe injury or death Hazards which if not avoided COULD result in severe injury or death Hazards or unsafe practices which if not avoided MAY result in injury or property damage Read and unders...

Страница 4: ...provided by the unit Failure to observe these warnings could result in severe injury or death Electric shock hazard Do not operate with the case open Before opening the case remove the test leads or j...

Страница 5: ...Do not change the measurement function while the test leads are connected to a component or circuit Failure to observe this precaution may result in injury and can damage the unit Statement of Confor...

Страница 6: ...r 18 Relative measurement is enabled 19 M Mega 106 20 Ohms 21 Hz Hertz 22 k kilo 103 23 A Amps 24 V Volts 25 m milli 10 3 26 O L Overload indicator 27 Negative polarity indicator 28 AC measurement is...

Страница 7: ...y give higher sensitivity Reading may be zero when sensitivity is too low Electrical noise may cause unstable readings Hold Button HOLD Press momentarily to hold the present value on the display will...

Страница 8: ...RMS reading method uses internal circuitry to read the true RMS value This method is accurate within the speci ed crest factor limitations whether the input signal is a pure sine wave square wave tria...

Страница 9: ...eading from the circuit or component to be tested Settings Table To measure this value Set the Selector switch to this symbol This icon will appear on the display Connect red lead to Connect black lea...

Страница 10: ...wire in the jaw for highest accuracy Clamp Around Line Splitter Notes The Greenlee 93 30 Line Splitter is divided One section renders amps the other renders amps multiplied by 10 Close the jaw comple...

Страница 11: ...11 CM 450 Voltage Typical Measurements Resistance www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30...

Страница 12: ...12 Forward Bias Reverse Bias Typical Measurements Continuity Diode Test www calcert com sales calcert com 1 800 544 2843 0 5 10 15 20 25 30...

Страница 13: ...equency Range 0 40 to 40 00 A 1 5 0 08 A 50 to 60 Hz 40 0 to 400 0 A 1 5 0 8 A 400 to 600 A 1 5 8 A Adjacent conductor in uence 0 05 A A Add 2 if the conductor is beyond the position error marks Multi...

Страница 14: ...ction Sensitivity Sine RMS Range 400 0 mV 350 mV 10 Hz to 2 kHz 4 000 V 1 V 10 Hz to 5 kHz 40 00 V 32 V 10 Hz to 100 kHz 400 0 V 100 V 10 Hz to 10 kHz 600 V 500 V 10 Hz to 5 kHz 400 0 A 60 A 40 Hz to...

Страница 15: ...measurement categories are explained in more detail by the International Electrotechnical Commission refer to either of their publications IEC 61010 1 or IEC 60664 Measurement Category I Signal level...

Страница 16: ...case open Before opening the case remove the test leads or jaw from the circuit and shut off the unit Failure to observe these warnings could result in severe injury or death 1 Disconnect the unit fro...

Страница 17: ...Avant d ouvrir le bo tier retirer les ls d essai ou la pince du circuit et mettre l appareil hors tension L inobservation de ces consignes pourrait entra ner des blessures graves voire mortelles 1 D b...

Отзывы: