   UM-35ACI Data Sheet (UM03)

Functional Diagram

Connector Pinouts

Component Layout

Signal Conditioning Components

Calibration Procedure

UM-Series are connectable using the TB-KIT screw terminal
blocks provided with the meter. For greatest convenience, order
a Texmate Push-On screw terminal connector. Alternatively, a
pcb edge connector can be used.(see connector options)

Pins 1 & 2 - Display Test: All numeric display segments will light
up when this pin is connected to the H/T Common Pin. A Texmate
TB-KIT Screw Terminal Clip can be used to access the Display Test
function.
Pins 3, 4, C & D - H/T Common Pin: The  Hold and Display
Test pins have to be connected to this pin to activate their
respective functions.
Pins A & B - Hold Reading: If this Pin is left unconnected, the
meter will operate in a free-running mode. When this pin is con-
nected to the H/T Common pin, the meter will latch up. A/D con-
versions will continue, but the display will not be updated until
Pins A & B are disconnected from the H/T Common pin. If this
function is to be accessed through a Texmate TB-KIT Screw
Terminal Clip, then jumper  J5 will have to be opened to dis-
connect the Test function. If both hold and test functions need
to be accessed, a PCB edge connector (part no. CN-L15)
should be used.

Pins 6, 7, F & H - Signal High Input: Signal High input for the
meter. Two factory installed ranges for direct connection to a 1A
or 5A CT (Current Transformer).
Pins 8, 9, J & K - Signal Low Input: Signal low input of the A/D
Converter.

Pins 11 & M - AC1 - Live AC Power Input:
Pins 12 & N - AC2 - 110/220V AC Power Select:
Pins 13 & P - AC3 - 110/220V AC Power Select:
Pins 14 & R - AC4 - Neutral AC Power Input:

A

1

2

3

4

5

J5

6

7

8

9 10 11 12 13 14 15

B

C

D

E

F

H

K

J

L M N

P R

S

PCB

A

B

C

D

E

TST

H/T

I/P

LO

I/P

HI

1

2

3

4

6

7

8

9

11

12

13

14

A

B

C

D

E

HLD

H/T

I/P

LO

I/P

HI

A
B

C

D

F
H

J

K

M

N

P

R

Span Pot

1.XXX
1X.XX
1XX.X
1XXX.

Decimal
Select
Header

Decimal Select Header

Cut loop to turn OFF last digit for 199 display

0.8" Optional Display

0.56" Display

Common

Input HI

Input LO

Hold

Test

J5

TST

H/T

HLD

SIG

LO

SIG

HI

AC Neutral

AC Line

240V

120V

AC1

AC2

AC3

AC4

GND

Input HI

Input LO

Hold

470K

5.1K

5.1K

0.22 F

Test

12 Bit 

Dual Slope 

A to D and

Display Driver

+ 5 V — 5 V

GND

To Display

Span Pot

Zero Pot

Zero Pot

1.XXX
1X.XX
1XX.X
1XXX.

5K

9K53

47K

— 5 V

+ 5 V

+ 1.25V

Zero Pot

40K2

J3

Input

Protection

Circuit

47K

1M

Span

Pot

+ 5 V

— 5 V

Differential

to Single

 Ended

Conversion

+ 5 V — 5 V

AC to DC

Conversion

Circuit

1Amp (0.1‰)

or

5Amp (0.02‰)

Shunt

Filter

Circuit

Rectifier

and

Regulator

Circuit

GND

-5 V DC

+5 V DC

!

WARNING

:

AC and DC input signals and power supply

voltages can be hazardous. Do Not connect live wires to
screw terminal plugs, and do not insert, remove or handle
screw terminal plugs with live wires connected.

SPAN Potentiometer (Pot)

The 15 turn SPAN pot is always on the right side
(as viewed from the front of the meter). Typical
adjustment is 100% of the input signal range.

ZERO Potentiometer (Pot)

The ZERO pot is to the left of the SPAN pot (as
viewed from the front of the meter). Typically it
enables the displayed readings to be offset
±500 counts.

1. Apply an input of 0 amps AC to the meter by shorting the

inputs. Adjust the Zero pot until the meter reads 000.

2. Connect the secondary of the current transformer (CT) to

the meter inputs, and apply a known current. For high cur-
rent CTs, a known AC current, proportionate to the CT out-
put, should be used for calibration.

3. Adjust the Span Pot until the meter displays the required

reading for the current being applied.

4. The UM-35ACI is now calibrated and ready for use.

(Whenever a new range is selected, re-calibration is required to
meet the specified accuracy).

1A or 5A

Current Shunt

Zero Adjust

Potentiometer

J5

Input High

AC Amps

Input Low

Span Adjust

Potentiometer

AC Neutral

AC Line

240V

120V

A

B

C

D

E

A
C
1

A
C
2

A
C
3

A
C
4

N/C

A

B

C

D

E

A
C
1

A
C
2

A
C
3

A
C
4

AC2 & 3

Joined

AC2 & 3 Joined

A

B

C

D

E

A
C
1

A
C
2

A
C
3

A
C
4

    L    M    N    P   R   S

   10  11   12  13  14  15    

    L    M    N    P   R   S

   10  11   12  13  14  15    

    L    M    N    P   R   S

   10  11   12  13  14  15    

See below for

connections

Отзывы: