45
Query
example
:SYST:CONF:BEEP?
>1
Return the setting of buzzer.
4-11-3. :SYSTem:CONFigure:BLEeder[:STATe]
Set
Query
Description
Sets or queries the status of the bleeder resistor (F-09).
Syntax
:SYSTem:CONFigure:BLEeder[:STATe] {<NR1> | OFF | ON
|AUTO }
Query Syntax :SYSTem:CONFigure:BLEeder[:STATe]?
Parameter
0 / OFF
Turns the bleeder resistor off.
1 / ON
Turns the bleeder resistor on.
2 / AUTO Turn the AUTO mode on.
Return
parameter
<NR1>
Returns bleeder resistor status.
Example
:SYST:CONF:BLE ON
Turns the bleeder resistor on.
Query
example
:SYST:CONF:BLE?
>1
Returns bleeder resistor status.
4-11-4. :SYSTem:CONFigure:CURRent:CONTrol
Set
Query
Description
Sets or queries the CC control mode (local control (panel),
external voltage control, external resistance control). (F-91)
This setting is applied only after the unit is reset.
Syntax
:SYSTem:CONFigure:CURRent:CONTrol
{<NR1> | NONE | VOLTage | RRISing | RFALling}
Query Syntax :SYSTem:CONFigure:CURRent:CONTrol?
Parameter
<NR1>
Description
0 / NONE
Local (Panel) control
1 / VOLTage External voltage control
2 / RRISing External resistance control;
10kΩ = Io max, 0kΩ = Io min.
3 / RFALling External resistance control;
10kΩ = Io min, 0kΩ = Io max.
Example
:SYST:CONF:CURR:CONT 0
Sets the setting state of the CC control mode.
Query
example
:SYST:CONF:CURR:CONT?
>0
Return the setting state of the CC control mode.