Figure 25. Fault Status tab
Table 6. Fault Status Indicators
Fault Indicator
Description
I2C_ERR
An error response on the access from I2C during access to the slave
OWI_ERR
An error response on the access from OWI during access to the slave
TF_ERR
Error response on the access from Trace FIFO
SYNC <320BPS
SYNC field data rate under 320 bits per second
SYNC >9600BPS
SYNC field data rate over 9600 bits per second
SYNC STOP Short
SYNC stop bit too short
CMD STOP Bad
Incorrect CMD stop bit value
CMD STOP Short
CMD stop bit too short
Data STOP Bad
Incorrect DATA stop bit value
Data STOP Short
DATA stop bit too short
Data STOP Override
DATA Field slave transmit value overdriven to dominant value during stop bit transmit
SYNC Tolerance Err
Consecutive bits in the sync field are different by more than ±25% tolerance