Appendix G
DLPU030B – June 2015 – Revised July 2015
DLP NIRscan Nano Command Description
G.1
Command Handler Supported Commands
The DLP NIRscan Nano Command Handler supports a set of commands described in
Table G-1. DLP NIRscan Nano Supported Commands
COMMUNICATIONS INTERFACE
USB
BLUETOOTH
READ /
DESCRIPTION
WRITE
OPCODE
CHARACTERISTI
SERVICE NAME
SERVICE UUID
C UUID
BYTE 1
BYTE 2
READ
Perform file checksum
0x00
0x15
WRITE
Write file data
0x00
0x25
WRITE
Write file size
0x00
0x2C
READ
Read file size
0x00
0x2D
READ
Read file data
0x00
0x2E
WRITE
Update Tiva firmware
0x00
0x2F
READ
EEPROM test
0x01
0x01
READ
Detector board test
0x01
0x02
READ
Battery charger test
0x01
0x03
READ
SDRAM test
0x01
0x04
WRITE
DLP Controller power up test
0x01
0x05
READ
Temperature sensor test
0x01
0x06
READ
Humidity sensor test
0x01
0x07
WRITE
Bluetooth test
0x01
0x08
READ
microSD Card test
0x01
0x09
READ
LED test
0x01
0x0B
READ
Read button test
0x01
0x0C
WRITE
Write button test
0x01
0x0D
FACTORY USE ONLY: Write
EEPROM Calibration
Coefficient, Scan
WRITE
0x01
0x0E
Configuration, and Reference
Calibration Coefficients
Versions
(1)
0x2829, 0x2A24,
Device Information
READ
Read version information
0x02
0x16
0x180A
0x2A26, 0x2A27,
Service
0x2A28
WRITE
Write patterns in SDRAM
0x02
0x17
GATT Scan Data
0x53455206-444C- 0x4348411D-444C-
WRITE
Start scan
0x02
0x18
Information Service
5020-4E49-
5020-4E49-
52204E616E6F
52204E616E6F
(1)
Commands highlighted in red are for factory calibration. Erasing the factory calibration data will render a unit non-functional and
will need to be recalibrated.
58
DLP NIRscan Nano Command Description
DLPU030B – June 2015 – Revised July 2015
Copyright © 2015, Texas Instruments Incorporated