20
EN
The set-point temperature and humidity values can
be modified using the
and
keys at any
time.
When one of the
or
, keys are pressed,
the temperature digits are replaced with the set-
point value in negative to indicate access to “set-
point modification” mode; press the key again to
change the values.
After 3” of inactivity of the
and
, keys,
you quit “set-point modification” mode, the tempera-
ture goes back to being displayed but not in nega-
tive and the new set-point temperature value be-
comes operational.
The same procedure must be carried out to modify
the humidity set-point, after selecting it using the
and
keys.
When the temperature set-point value chang-
es, the system changes the humidity value
automatically, and vice versa, in order to
maintain usage conditions that do not jeop-
ardise the health of the product user.
Thresholds of temperature and humidity setting are
as follows:
TEMPERATURE
MAX HUMIDITY
SETTABLE
45°C
100%
50°C
60%
55°C
50%
60°C
40%
110°C
30%
HUMIDITY
MAX.
SETTABLE
TEMPERATURE
30%
110°C
40%
60°C
50%
55°C
60%
50°C
100%
45°C
3sec.