2 The routing within the network infrastructure is not working. Check whether
too many terminal devices are logged into the access point.
The WiFi data logger is displaying the error code E41, what can I do?
The WiFi data logger cannot obtain any current time from a time server
(pool.ntp.org).
1
The access point (WLAN router) has no connection to the internet.
Check the access point's internet connection.
2
The NTP port (123/UDP) of the access point (WLAN router) is not open.
Check whether the NTP port (123/UDP) is opened.
The WiFi data logger is displaying the error code E51, what can I do?
The WiFi data logger was not able to connect to the Testo Cloud.
1
If the WiFi data logger has already been connected to the Testo Cloud
and this connection is suddenly no longer possible: The Testo Cloud
servers are not currently accessible. The servers will be monitored and
should be accessible again within a few hours.
2
If the WiFi data logger has not yet been connected to the Testo Cloud:
The TCP ports (1883 or 8883) of the access point (WLAN router) are not
open. Check whether the TCP ports (1883 or 8883) are open in both
directions.
The WiFi data logger is displaying error code E52, what can I do?
The data logger could not log into the Cloud because it is already logged into
another account. Please log the data logger out of the existing account first.
The WiFi data logger is displaying the error code E63, what can I do?
The WiFi data logger could not send any data to the Testo Cloud.
1
The internet connection was interrupted during the transmission. Check
whether there is a stable connection from the WiFi data logger to the
access point (WLAN router). Check the access point's internet
connection. The data will be transferred during the next communication
cycle. Alternatively: Initiate data transmission manually by pressing the
control key on the WiFi data logger.
2
The Testo Cloud server was not able to process the request for data
storage. The servers will be monitored and should be accessible again
within a few hours.
The WiFi data logger is displaying the error code E75, what can I do?
A firmware update for the WiFi data logger failed.
The internet connection was interrupted during the transmission or the data was
not received intact by the WiFi data logger for other reasons. Check whether there
is a stable connection from the WiFi data logger to the access point (WLAN
router). Check the access point's internet connection. The data will be transferred