background image

Adjustment

Adjustment in 

environmental chamber

Depending on application and exposure to dirt an annual

inspection in Testo`s calibration laboratory is

recommended.

Adjustment can be carried out in an environmental

chamber or at Testo.
ISO calibration certificates for humidity and temperature

are available.

In the environmental chamber, you can adjust the heated

compact transmitter using the adjustment buttons.

Carry out adjustment as follows: 

Place transmitter in the environmental chamber and set

adjustment point 11.3 %RH at 25 °C in the

environmental chamber.

Wait for adaptation time to elapse (at least one hour).

Release cap sleeve at bottom end. The adjustment

buttons are visible.

Press S1 for adjustment point 11.3 %RH.

Carry out steps 

➀ 

to 

➂ 

for the adjustment point 

75.3 %RH. Press button S2 for adjustment point 75.3

%RH.

Screw on cap sleeve following adjustment.

The heated compact transmitter cannot be adjusted using

the control and adjustment set.

7

S1

S2

Содержание Hygrotest 6337 9742

Страница 1: ...Bedienungsanleitung Hygrotest 6337 9742 Instruction manual en de...

Страница 2: ...al Probleme auftreten die Sie nicht selbst beheben k nnen wenden Sie sich bitte an unseren Kundenservice oder Ihren H ndler Wir bem hen uns schnelle und kompetente Hilfe zu leisten damit Ihnen lange A...

Страница 3: ...ahmen nicht getroffen werden Lesen Sie Warnhinweise besonders aufmerksam und treffen Sie die genannten Vorsichtsma nahmen um Gefahren zu vermeiden Hinweise auf Sonderf lle oder Besonderheiten im Umgan...

Страница 4: ...e niemals Gewalt an Temperaturangaben auf Sonden F hlern beziehen sich nur auf den Messbereich der Sensorik Setzen Sie Handgriffe und Zuleitungen keinen Temperaturen ber 70 C aus wenn diese nicht ausd...

Страница 5: ...besonders zum Einsatz bei hohen Feuchten und schnell wechselnden Temperaturen Hier kann Feuchtigkeit auf der Oberfl che der Bauteile durch die verz gerte Temperaturanpassung der Bauteile kondensieren...

Страница 6: ...nnungssignalen zu beachten Wird der beheizte Feuchtesensor bei h heren Tempera turen als Umgebuns und Lagertemperatur und sehr hoher Feuchte hohe Taupunkte eingebaut ist eine An gleichzeit zu beachten...

Страница 7: ...abgleichen Abgleich durchf hren Messwertumformer in den Klimaschrank bringen und Abgleichpunkt 11 3 rF bei 25 C im Klimaschrank einstellen Angleichzeit abwarten mindesten eine Stunde berwurfh lse am u...

Страница 8: ...Sinterkappen mit reinem Alkohol oder des tillierten Wasser reinigen oder Kappe mit Druckluft aus blasen von innen nach au en Sensor nicht ber hren und nicht ansto en Sensor wechseln Defekte Sensoren...

Страница 9: ...00 4 Leiter durchgeschleift Zul ssige B rde Spannungsausgang RL 10 k Stromversorgung 18 28 V DC Zul ssige Lager und Transporttemperatur 20 80 C Abmessungen F hlerl nge 196 mm Sondenrohr 21 mm Beschrei...

Страница 10: ...Notizen...

Страница 11: ...Bedienungsanleitung Instruction manual en de Hygrotest 6337 9742...

Страница 12: ...ich you cannot rectify yourself please consult our service department or your dealer We will endeavour to provide fast and competent assistance to avoid lengthy down times Copyright This documentation...

Страница 13: ...not carried out Read all the warnings carefully and carry out all the specified safety measures to avoid danger Notes on special cases and peculiarities in the handling of your unit are indicated by...

Страница 14: ...re Force should never be applied The temperature data for the sensors probes only refer to the sensor measuring range Never subject handles or pipes to temperatures greater than 70 C unless they are e...

Страница 15: ...y suitable for applications with high moisture levels and quickly changing temperatures Humidity can condense on the surface of the components due to the delayed temperature adaptation of the componen...

Страница 16: ...account If the heated humidity sensor is installed at higher temperatures ambient and storage temperature and a very high humidity level high dew points an adaptation time should be adhered to The ada...

Страница 17: ...tter using the adjustment buttons Carry out adjustment as follows Place transmitter in the environmental chamber and set adjustment point 11 3 RH at 25 C in the environmental chamber Wait for adaptati...

Страница 18: ...an dirty sintered caps with pure alcohol or destilled water or purge cap with compressed air from inside outwards Do not touch sensor or subject to impact Changing the sensor Defective sensors can be...

Страница 19: ...Temperature Pt100 4 wire looped through Maximum load Voltage output RL 10 k Power supply 18 to 28 V DC Maximum storage and transport temperature 20 to 80 C Dimensions Probe length 196 mm Probe pipe 2...

Страница 20: ...testo AG Postfach 11 40 79849 Lenzkirch Testo Stra e 1 79853 Lenzkirch Telefon 07653 681 0 Fax 07653 681 100 E Mail info testo de Internet http www testo com 0973 0618 T wh PC_qxd 0603...

Отзывы: