Condensate trap
Changing filters
The filter must be exchanged if dirt particles can be seen.
To do this, unlock the condensate trap and remove from housing.
Remove the filter and replace with a new one. Order spare filter
(Part no. 0554.0040).
Only this filter should be used.
A tightness test should be carried out every time a filter is
changed (See page 3).
The max. level in the condensate trap should not be exceeded.
Remove the emptying plugs to empty the condensate trap.
Remove empty/defect rechargeable battery or empty battery from
the battery compartment and replace with new rechargeable
batteries or batteries.
Warning
Observe the correct polarity of the rechargeable
batteries/batteries
Switch off analyser before changing battery or
rechargeable battery.
Complete change in 1 -2 minutes or plug in mains
unit.
The set date/time may otherwise be lost.
Empty the condensate trap
before changing filter.
14
Maintenance
Changing rechargeable batteries or batteries
Filter
Max. level
indicator
Condensate
emptying
1
2
3
4
m a x
max