6 Using the product
14
Function
Setting options/comments
calibration
humidity
yes
or
no
yes
selected:
Humidity values are aligned to the alignment
points 11.3% RH and 75.3% RH and saved.
To adjust to the alignment points, use
the testo
Control and alignment set
for humidity sensors
(0554 0660).
> Cancel configuration menu: press and hold down
[ ]
until the
configuration mode
display switches to the measurement
view.
6.2.
Measuring
✓
Instrument is switched on.
- Display of the current CO reading.
Switching to humidity, temperature and dew point value is
only possible if a humidity/temperature module is
connected.
> Display humidity and temperature value: press
[ ]
briefly.
> Display dew point: press
[ ]
briefly.
Delta measurement
✓
Instrument is in measurement view.
> Call up Delta measurement menu: press
[ ]
until the display
changes.
- Delta measurement is carried out (
[
]
lights up). The current
readings are zeroed as a reference.
> To cancel the function: press and hold down
[ ]
until the
display switches to the measurement view.