background image

12

3.

Product description

3.1 Overview

Overview



Contact plates



Handle

‘

Connection cable

3.2 Humidity measurement

procedure

The non-destructive stray field measurement uses the
ability of water molecules to dampen and thus change
electromagnetic fields. The electric field penetrates the
material via the contact plates and creates a measuring
field with a depth of approx. 5 cm.
The following factors can influence the measurement result:

Factors

Ideal condition

Measurement depth

Material thickness > 5 cm (> 2“). Caution: The upper

layers of the material influence the measurement result

more than the deeper layers.

Material surface 

As level as possible as contact plates should lay flat

against it.

Material properties

As homogeneous as possible with no air pockets.

Moisture distribution

As even as possible.

Metals and electric fields

None, if possible.

3. Product description

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание 0636 6160

Страница 1: ...n Meaning Comments Warning advice Warning Read warning advice carefully and take the precautionary measures indicated Serious physical injury could occur if you do not take the precautionary measures...

Страница 2: ...Do not expose handles and feed lines to temperatures in excess of 70 C unless they are expressly permitted for higher temperatures Temperatures given on probes sensors relate only to the measuring ra...

Страница 3: ...such as CM method or Darr W ge dry and weigh method The probe is not calibratable For connection to testo 635 1 Art no 0560 6351 and testo 635 2 Art no 0560 6352 The product must not be used in the f...

Страница 4: ...of approx 5 cm The following factors can influence the measurement result Factors Ideal condition Measurement depth Material thickness 5 cm 2 Caution The upper layers of the material influence the mea...

Страница 5: ...to contact surface Slowly increase contact pressure to 1 to 3 kg 2 2 to 4 4lb until a stable value is shown In order to be able to evaluate a moisture course several measurements at different positio...

Страница 6: ...stomer service 6 Technical data Characteristic Values Parameters Water content in weight percent of dry mass Measuring ranges Woods 50 Building materials 20 Resolution 0 1 Probe Contact plate integrat...

Страница 7: ...limber 9 3 by weight Chipboard 8 by weight Cement screed 3 by weight Anhydrite screed 0 5 by weight Concrete 2 2 by weight Solid brick 1 by weight High insulating brick 2 5 by weight Limestone 1 3 by...