10 Tips and assistance 

19 

 

10 Tips and assistance 

 

10.1  Questions and answers 

Question 

Possible cause 

Possible solution 

Zero point is unstable 

Contamination of the 
sensor after a prolonged 
period of non-use 

Leave the instrument 
switched on until the 
zero point has stabilized. 

Instrument does not 
switch to measuring 
mode (remains in warm-
up phase) 

Battery voltage too low 

Charge the instrument. 

 

 

10.2  Error codes 

Error code 

Error 

E001 

Memory error 

E002 

Sensor error 

 

The error code is shown on the display. 

 

10.2.1  Hard Reset 

If you encounter any problems with the firmware, press and hold the 

ON/OFF

 key for a long time (4 sec.) to perform a reset. 

 

 

10.3  Accessories and spare parts 

Description 

Order no. 

Transport bag 

0590 0018 

Mains unit from Testo 

0664 1107 

 

 

11 Support 

You can find up-to-date information on products, downloads and links to contact 
addresses for support queries on the Testo website at: www.testo.com. 

 

If you have any questions please contact your local dealer or the Testo 
Customer Service. You can find contact details on the back of this document or 
online at 

www.testo.com/service-contact.

 

 

 

Содержание 0560 3163

Страница 1: ...testo 316 1 EX gas leak detector 0560 3163 Instruction manual...

Страница 2: ......

Страница 3: ...tion 8 6 2 1 Establishing a Bluetooth connection to the testo Smart App 8 7 Using the product 9 7 1 Controls 9 7 1 1 Implementing settings 11 7 2 Carrying out a function test 13 7 3 Carrying out gas d...

Страница 4: ......

Страница 5: ...Do not operate the instrument if there are signs of damage on the housing mains unit or connected cables Dangers may also arise from objects to be measured or the measuring environment Always comply...

Страница 6: ...tteries in accordance with the valid legal specifications At the end of its useful life deliver the product to the separate collection point for electric and electronic devices observe local regulatio...

Страница 7: ...lammable substance in the air has a lower explosive limit LEL and an upper explosive limit UEL The air gas mixture is flammable anywhere between these two limits potentially leading to an explosion cr...

Страница 8: ...play 5 Control keys 6 Charging socket for rechargeable battery Symbol explanation Do not allow children under 6 years of age to play with batteries Do not throw batteries in the trash Do not charge ba...

Страница 9: ...on the longer this additional warming up phase should be Please note that the instrument switches itself off by default after 10 min of inactivity 1 Press and hold down 1 sec the On Off key Warm up ph...

Страница 10: ...ing is confirmed by AOFF ON or AOFF OFF 6 2 Establishing a Bluetooth connection The instrument can be connected to the testo Smart App via Bluetooth connection The gas leak detector is switched on 6 2...

Страница 11: ...The app is installed on the smartphone and connected to the instrument via Bluetooth Settings and controls are primarily carried out on the instrument and are transferred to the app To a lesser extent...

Страница 12: ...measurement data report Searching Favourite Deleting Further information Display report Multiple selection The Main menu can be accessed via the icon at top left To exit the main menu select a menu o...

Страница 13: ...ng and holding the key 1 sec Adjustable functions Ensure correct settings all settings are transferred immediately There is no Cancel function Function Setting options comments On Off long press Switc...

Страница 14: ...of the alarm sound increases as the concentration increases Unit long press Switch between PPM and LEL Measuring mode MODE 1 for gas detection MODE 2 for locating leaks Max zeroing long press Zero max...

Страница 15: ...rer Testing natural gas lines or hydrogen lines Methane main component of natural gas or hydrogen are lighter than air detection should be carried out above the pipe suspected leak Testing propane and...

Страница 16: ...falling concentration values are output At the beginning two bar segments are displayed in the middle At the same time the sensor LED flashes in a colour corresponding to the concentration value and a...

Страница 17: ...k it is advisable in the case of these larger leaks to switch to MODE 2 in order to receive feedback via audio and flashing signals If the warning threshold 100 ppm is exceeded the sensor LED and disp...

Страница 18: ...using max 0 After the measurement 1 Ventilate the sensor thoroughly after each use To do this place the instrument in fresh air for approx 2 minutes before using it again 8 Maintaining the product 8...

Страница 19: ...r inspected by an authorized service centre every year 8 3 Cleaning the sensor Tobacco smoke dirty air oils greases silicones and evaporating liquids or gases can leave deposits on the sensor surface...

Страница 20: ...se time 2 sec Leak alarm 3 colour LED on sensor head 3 colour display backlighting Acoustic App Operating temperature 5 to 50 C Operating humidity 0 to 80 RH Operating altitude 2000 m Storage temperat...

Страница 21: ...ensor error The error code is shown on the display 10 2 1 Hard Reset 1 If you encounter any problems with the firmware press and hold the ON OFF key for a long time 4 sec to perform a reset 10 3 Acces...

Страница 22: ...Testo SE Co KGaA Celsiusstr 1 79822 Titisee Neustadt Germany Phone 49 0 7653 681 0 Email info testo de www testo com 0970 3263 en 01 08 2022...

Отзывы: