Vibrating Table
2.0231 / 2.0268 / 2.0269
2.0131 / 2.0233 / 2.0234
2.0271
15
3.
Placing the Vibrating Table into operation
The device may only be operated in dry rooms!
The following limit values apply:
Permissible temperature:
from 5 °C to +40 °C
Operating humidity range:
from 20 % to 75 % r.H.
Be sure to set up the Vibrating Table on a level surface which can satisfactorily support
its weight and the vibration forces.
Carry out a short functional test by connecting the mains plug to the mains supply.
Caution
We strongly recommend that you use a residual-current-operated
circuit-breaker (RCCB; also known as earth-leakage circuit-breaker,
ELCB) with this Vibrating Table. This device should have a fault-cur-
rent tripping rating of 30 mA.
4.
Operation
Danger
Be aware that incorrect inputs to the system can lead to malfunctions,
breakdown of the system, major material damage or to danger to the
operating personnel.
Standard vibrating tables without magnetic clamping device
Place a filled concrete mould on the centre of the vibratory plate of the table, and hold it
securely by hand.
Start the vibration function by using the pedal switch / manual switch or time switch.
During the vibration process, the operator must hold the concrete mould securely by
hand.
After the desired time has lapsed, or after it has become visually apparent that the desired
results have been achieved, stop the vibration function by using the pedal / manual switch
or time switch again.
Attention
Remove overflowing material with a wet cloth. Do not keep the appli-
ance under water. It is only protected against splash water!
Set the desired vibration time (from a minimum of 1 second to a maximum of 9,999 sec-
onds)
Start the vibration function by using the START/STOP button.