TESTING 1.0290 Скачать руководство пользователя страница 13

 
 
Electronic 
Blaine Air-Permeability Apparatus 
1.0290/1.0290E 
 

 

13 

6. 

Troubleshooting 

 
This section describes a number of simple problems tht can be easily solved during work. 

 

Caution 

 
 

All maintenance, inspection, testing, and repair work on apparatus 
components or on the electrical system may be performed ONLY 
by sufficiently qualified personnel.  
 

 

PROBLEM 

CAUSE 

SOLUTION 

The system will 
not start. 

There is no power to the system. 
 
 
 

 

The pump or the electrical system 
is defective. 

 

The U-shaped tube is covered 
with condensation or dirt. 

Operate the main power switch 
correctly, check the power ca-
ble, fuse on the main switch 
and exchange it if necessary. 

 

Get in touch with the Supplier. 
 

 

Clean the U-shaped tube. 

The manometer 
fluid is not pulled 
in correctly. 
 

pump, valve is defective 
The pneumatic lines are loose or 
have leaks. 

 

The U-shaped tube is covered 
with condensation or dirt. 

Get in touch with the Supplier. 
 
 

 

Clean the U-shaped tube. 

The measured val-
ues are not cor-
rect.  

The constant for the apparatus is 
not correct. 

 

The calibration substance does 
not properly match the cement 
type. 

 

The time metering is not correct. 
The photoelectric barrier does not 
function. 
 

 

The measuring cell has not been 
properly attached, or it has not 
been turned correctly. There is a 
leak in the connection between 
the fittings. 

Correctly determine the con-
stant for the system. 

 

Choose the proper calibration 
substance to correctly match 
the cement type. 

 

Regulate the photoelectric bar-
riers on the potentiometer. 
Check the time metering. Clean 
the U-shaped tube. 

 

Check the glass cone  

 

Note 

 

If the device is to be sent for repair, the manometer liquid must be 
sucked off in order to avoid further damages to the device. 

 

Содержание 1.0290

Страница 1: ...Distribution of Systems for the Testing of Construction Materials Motzener Str 26b DE 12277 Berlin Tel 49 30 7109645 0 Fax 49 30 7109645 98 1 Operating Manual Electronic Blaine Air Permeability Appar...

Страница 2: ...ns 5 1 6 Acceptance of the product and transport 6 1 6 1 Acceptance of the product 6 1 6 2 Transport 7 1 7 Scope of delivery 7 1 8 Instructions for electrical connection 8 2 Characteristics of the app...

Страница 3: ...nition of technically qualified staff has been provided in IEC 364 The User must by all means observe the requirements and limit values as well as all safety instructions given in this Operating Manua...

Страница 4: ...with the following actions being taken It may therefore NOT be used under such conditions or in connection with the following actions Do not disassemble this system Do not try to repair it or to modif...

Страница 5: ...criteria Persons who are at least eighteen 18 years of age and Persons who have been instructed in the operation of the Blaine Apparatus and In the operation of the Blaine Apparatus the person operati...

Страница 6: ...e descriptions given in this Operating Manual all applicable safety directives guidelines and regulations and all general rules for workplace environments The Operator must ensure that the personnel w...

Страница 7: ...o the point at which it is to be operated Its weight is approx 4 kg After you have unpacked the device make a visual inspection to determine whether it was damaged during transport In cases of doubt i...

Страница 8: ...ystem against overcurrent This device must match the machine voltage and must be in accordance with the valid regulations The technical characteristics of this safety device must satisfy the reg ulati...

Страница 9: ...provided by a separate power cable with a two pole and earthing pin plug The Manufacturer tests the function and liquid tightness of the apparatus before delivery 2 2 Technical data Power rating 110...

Страница 10: ...the injector hose into the U shaped tube Make sure that you can see the hose in the U shaped tube and also make sure that the manometer liquid actually flows into the tube Fill the liquid up to the lo...

Страница 11: ...mps of the oil level indicators will now show consecutively green amber and red The READY lamp will now go off The red lamp will now show which means that the digital counter is set to zero The pump s...

Страница 12: ...pparatus from the power supply Use a brush or vacuum cleaner to remove loose dust on the apparatus If neces sary the apparatus can then be cleaned with a moist cloth Use a normal household cleaning ag...

Страница 13: ...ctly pump valve is defective The pneumatic lines are loose or have leaks The U shaped tube is covered with condensation or dirt Get in touch with the Supplier Clean the U shaped tube The measured val...

Страница 14: ...ase see the diagram below 7 Shutting down the system for lengthy periods If the Blaine Apparatus is scheduled to remain out of operation for a lengthy period of time please follow these steps Disconne...

Страница 15: ...llowing company TESTING Bluhm Feuerherdt GmbH This Operating Manual is provided only for the using company and the staff of this com pany The information in this Operating Manual may not be Reproduced...

Отзывы: